25
6.2.4 MONITORING screen
Monitoring window
Touch the “Monitoring” button to view EUT power input parameters, and to control test activity and
EUT power input in the event of EUT failure.
EUT Supply Voltage, EUT Supply Frequency
In case a CDN 3153-S63 is connected to the NSG 3150
, the “EUT Supply Voltage” field displays the
actual EUT voltage when the AC EUT input supply is connected and EUT power is switched “On”.
When the input supply is not connected and / or the EUT is switched off, these fields will display NA.
The EUT Supply frequency field shows the measured frequency of the EUT supply voltage.
Test Action at EUT Fail Input
Touch the “Test Action at EUT Fail Input” button (“Stop” in the example) to specify the test action
taken if the “EUT fail input” (on system Interface port) is activated.
When the button is set to “Stop” and the “EUT fail input” is activated, the test stops. The test can be
restarted by pressing the Start key on the front panel.
When the button is set to “Pause” and the “EUT fail input” is activated, the test goes into pause mode.
The test can be continued by pressing the “Start” key on the front panel. When the button is set to
“CONT.”, the test will continue even if the EUT fail.
EUT Power Supply at EUT Fail Input
Touch the “EUT Power Supply at EUT Fail Input” button (not shown on example) to specify the action
taken if an “EUT fail signal” is generated.
When the button is set to “On”, EUT power stays ON after the “EUT fail Input” is activated.
When the button is set to “Off”, EUT power shuts down when the “EUT fail Input” is activated.
EXIT
Touch the “Exit” button to return to the system window without saving changes.
EUT ON / OFF
This button displays the EUT input power status.
OK
Touch the
”Ok”
button to save changes and return to the system window.