5.14 Easy Calibration
radEAGLET
User Manual
Please place a suitable calibration source (
40
K or
137
Cs) in front of the detector before
proceeding.
After you let the
radEAGLET
know which calibration source is in front of it (Fig.
, p.
), it
will aquire data for a while and use it for calibration (Fig.
, p.
The acquired spectrum is shown, overlayed by the theoretical peak position of the selected
source and the current peak fit of the measured spectrum.
Figure 64:
Selecting a source for easy calibration
Figure 65:
Easy calibration in progress
A progress percentage is shown during the data collection, sometimes in several cycles,
until the peak position of the calibration source is within ±0.5 % of the theoretical position
(Fig.
, p.
).
68/140
Software 3.2.0 • Document 3.2.2o
© innoRIID GmbH • 2018-10-24