radEAGLET
User Manual
Welcome
Several models of the
radEAGLET
are available (see
, p.
for details), all of which
are covered in this document. The few cases where a feature of a certain model
differs from the main stream are marked in the text.
1.5
The
radEAGLET
The
radEAGLET
is a new generation radio-isotope identification device (RIID). It consists of
the following components:
• Scintillation detector using a sodium iodide NaI(Tl) crystal.
• Geiger-Müller detector
3
He Neutron detector
• Multi-Channel Analyzer (MCA) for spectral data readout of the scintillation detector
• Computational subsystem that includes TFT display, keyboard, status LEDs, vibrator
and speaker
• Pack for standard AA batteries
GPS Receiver
1.5.1
Detectors
Each component has a dedicated purpose. The scintillator is the primary detector of the
instrument and would be used for multiple purposes including pulse height analysis and dose
rates.
1.5.1.1 Scintillation Detector
The scintillation detector is used to collect the pulse height spectrum of the gamma photons
that interact with the scintillation crystal. The different radioisotopes each have specific
decay schemes and some emit gamma photons that can be analyzed and used to determine
the radiation source. (
Þ
Scintillation Detector)
1.5.1.2 Geiger-Müller Detector
The dose rate is determined by either the scintillation detector or the internal Geiger-Müller
tube. When the dose rate at the scintillator surface exceeds 200 µSv/h, the Geiger-Müller
tube will perform the dose rate measurement. This tube is suited for measuring dose rates
up to 1 Sv/h. (
Þ
Geiger-Müller detector)
If the Geiger-Müller detector kicks in, you are already in an extremely dangerous level
of radiation. You should increase distance and shielding between yourself and the
source. Additionally, you should restrict the time you stay within this field to an abso-
lute minimum.
© innoRIID GmbH • 2018-10-24
Software 3.2.0 • Document 3.2.2o
13/140
Summary of Contents for ORTEC radEAGLET
Page 18: ...18 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 30: ...30 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 38: ...38 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 42: ...42 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 90: ...90 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 116: ...116 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...
Page 132: ...132 140 Software 3 2 0 Document 3 2 2o innoRIID GmbH 2018 10 24...