122770 06
18-09-2007
35
2.2.1 Supplies
ATC-140/155/156/157
Isolated supply (refers to GND’)
+26V’ : 26.6V
±
1.8V
Non-isolated supplies (refer to GND)
+25 V : +24.6V
±
1.5V
VCC : +5V
±
0.25V
VEE : –20.2V
±
0.9V
+27V : 26.6V
±
1.6V
ATC-250/320/650
Isolated supply (refers to GND’)
+26V’ : 26.6V
±
1.8V
Non-isolated supplies (refers to GND)
+25 V : +24.6V
±
1.5V
VCC : +5V
±
0.25V
VEE : –20.2V
±
0.9V
+27V : 26.5V
±
1.5V
ATC-125
Refers to GND
VCC: 5V ±0,5V
24V: 24V ±1V
VEE:
-19V ±1V
Refers to ISOGND
ISO 26V: 26V ±1V
2.2.2. Oil
Problems
Spilling
If for any reason it is necessary to remove silicone oil spilling from the
interior or exterior of the instrument, it is advisable to use OS20 from
DOW CORNING.
OS20 is silicone oil with a viscosity low enough to dissolve the
silicone oil used for calibrating purposes.
Warning
Read the product information before use. Due to flammable
liquid, precautions might have to be taken.
Due to transport safety regulations, the liquid is not distributed by
AMETEK Denmark A/S, but is readily obtainable from your local
DOW CORNING distributor.
Summary of Contents for JOFRA ATC-125
Page 2: ...2 18 09 2007 122770 06 ...
Page 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Page 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Page 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Page 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Page 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...