52
18-09-2007
122770
06
The reference sensor is the one
placed in the right hole beneath in
the picture.
Disconnect the 2 sensor wires from
the Connection PCB.
When remounting a new sensor, it
must be placed in the hole together
with 0.5ml compound.
ATC-320/650
A/B:
Place the well unit upside down
with the wires pointing towards you.
The reference sensor is the one
placed to the right.
Cut the plastic strap holding
together the wires from the
reference sensor and the two
thermocouples with pink wires.
Note the position of the straps for
later assembly.
Remove the 3-mm screw, washer
and bracket holding the reference
sensor. Now it is possible to retract
the reference sensor from the unit.
Summary of Contents for JOFRA ATC-125
Page 2: ...2 18 09 2007 122770 06 ...
Page 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Page 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Page 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Page 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Page 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...