60
18-09-2007
122770
06
When connecting new Main PCB’s to old Power PCB’s they must be
connected with a cable 125232.
Power PCB
ATC-125
There is no DIP-Switch on the power PCB for the ATC-125 and the
Power PCB has no need for any further adjusting or configuration.
New Power PCB with new Main PCB
When version 2 Power PCBs are connected to version 2 Main PCBs,
the DIP-Switch Settings on the Power PCBs are as follows:
Model DIP-Switch
Setting
ATC-140 00010101
ATC-155/156/157 01000100
ATC-250
00010100
ATC-320 *
00000000
ATC-650 01000000
New Power
PCB
with old Main
PCB
When version 2 Power PCB’s are connected to version 1 Main
PCB’s, the DIP-Switch Settings on the Power PCB’s are as follows:
Model DIP-Switch
Setting
ATC-155/156/157 10001110
ATC-320* 00001010
ATC-650 10001010
When connecting new Power PCB’s with old Main PCB’s they must
be connected with a cable 125231.
* See section 2.3.3 for configuring the Power PCB, version 2.
2.3.2 Configuring the Power PCB for ATC-320 A/B
(version 1)
Remove the diode D14 from the Power PCB.
Summary of Contents for JOFRA ATC-125
Page 2: ...2 18 09 2007 122770 06 ...
Page 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Page 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Page 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Page 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Page 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...