38
18-09-2007
122770
06
When remounting the spill tray the
screws have to be cross tightened
with a force of approximately 1N/m.
For best result, tighten the screws
when the calibrator is heated to
80°C. This will soften the rubber
seal, making the junction vertically
waterproof.
After tightening the screws, replace
the silicone seal between the
handle and the spill tray. Put a thin
line of silicone between the spill
tray and the handle and smooth it
out with a wet finger. A god tip is to
recalibrate the instrument, before
the silicone is applied.
B.
Removal of Display and Keyboard Unit, all Models
(Exploded views, pos. 2).
Remove the grid plate by
unscrewing the 3 screws (all
models except ATC-140/250 A/B).
Remove the protection shield if
present (only in the early models of
ATC-320/650 A/B).
Summary of Contents for JOFRA ATC-125
Page 2: ...2 18 09 2007 122770 06 ...
Page 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Page 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Page 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Page 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Page 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...