6-2 | Dycor CG1000-RTP Oxygen Analyzer
System Messages
Calibration Aborted
Calibration has been aborted by the user.
Calibration Required
A regular calibration is required.
After a thermal calibration is performed you must always perform a regular
calibration.
During a thermal calibration the cell temperature is changed. A
system calibration cannot be performed for at least one hour. The
system calibration will determine the constants needed for the
Nernst equation.
If you see this message, select the
Calibrate Key
and perform a calibration. This message will not clear from the dis-
play by removing or restoring power to the analyzer.
Cal Verify Aborted
The verification was successfully aborted by the system.
Primary Calibration
The system is performing a Primary Calibration. The message con-
tinues to be displayed during the one-hour recovery period after the
thermal calibration.
System Calibrating
System calibration is in progress.
System Verifying Cal
System verification is in progress.