Controller / User Interface
|
4-5
System Serial #
View your system serial # and manufacturing #. Service personnel may ask for
this information.
1. From the
Setup
menu, select
System Serial #
.
2. Select
System Serial #
or
Manufacturing #
.
3. The system displays the system serial # or manufacturing #.
Primary Calibration
Perform a primary calibration on the zirconium oxide oxygen sensor.
Always attempt to perform a regular calibration before resorting to a
primary calibration for all cases other than when you install a new cell.
A primary calibration should only be done in the following cases:
•
when you replace your zirconium oxide oxygen cell;
•
when your system fails a regular calibration due to a deteriorating cell.
Regular calibration isn’t working
• Check to see if your calibration gas values match the values written on
the gas cylinders.
• Ensure that the correct flow of gases has been set and that the gases
are allotted enough time to flow through the sensor.
If these checks have been done and the system still can’t be properly cali-
brated you should do a primary calibration.
Performing a primary calibration
Before starting the primary calibration, make sure you have properly
entered your calibration gas values using the
Cal Gas Values
option
from the
Calibrate
key.
1. From the
Setup
key, select
Primary Calib
.
2. At this point, the system works as it does for a regular calibration.
A thermal calibration is not a
regular calibration.
NOTE
Summary of Contents for Dycor CG1000-RTP
Page 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Page 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...