Installation & Start-Up
|
3-11
System Start-Up
1.
Plug the power cord into the power entry module on the back of the
analyzer.
2.
Plug the other end of the power cord into the appropriate Ac power
outlet.
3.
Turn on the analyzer by pressing on the side of the power switch
marked 1. The power switch is located on the back of the analyzer.
4.
Adjust the flow of the sample gas to the desired level (see Chapter 2 -
Installation for flow requirements).
5.
Allow the analyzer to warm-up for at least one hour.
6.
set up your alarms, analog outputs, and serial communications.
7.
Perform a regular calibration.
8.
Read the sample gas.
Password Restrictions
If there is a password restriction, you must enter the password before you
can access menu functions that can change system settings.
Display
You can define which information to place on the first three lines of the
display using the
Display
option from the
Setup
key. You can also use the
Display
menu option to place your own text message on one of the first
three lines.
The last display line is reserved for system and error messages.
If an alarm condition exists, alarm information in the form of an Up
↑
or
Down
↓
arrow (corresponding to a high or low alarm) will appear next to
the process reading on that display line.
Auto Ranging Feature
The CG1000-RTP analyzer includes an auto ranging feature. If the oxygen
reading falls below 0.5% O
2
, the display switches to the ppm scale. Other-
wise, the system always displays readings using the percent scale.
NOTE
Summary of Contents for Dycor CG1000-RTP
Page 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Page 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Page 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...