IOM (Safety) - 

Installation, Operation and Maintenance

                      Optical Oxygen Analyzer

The 

OXYvisor

 measures from trace level (ppmv(g) / ppb(aq)) oxygen to percent level oxygen in both the gas and liquid phases.  It’s intended for 

indoor and outdoor use, as a continuous analytical measuring device, in manufacturing and industrial processes, along with laboratories. 

 

The analyzer is either pipe or wall mounted in the proximity of the measuring point.  The analyzer has a local display and can be operated (configuration, 

calibration, etc.) with the HMI via the “through-the-glass,” keypad or via the 

OXYvisor

 software on a PC.  The analyzer connects to a fiber optic 

cable, via a fiber optic SMA connection located in the easily accessible junction box.  The fiber optic cable links the analyzer to the process sensor, 

which is mounted directly into the process or within an extractive  sample system.  The analyzer is fully compatible with all of the Barben Optical Sensors 
(BOSx) which are sold separately or included with integrated Sample Calibration Panels (SCP).  

The purpose of this abridged manual is to specifically inform the user of all the pertinent and required information, as it pertains to safety to install and 

operate this product.  The product is to be used for its intended purposes, as detailed below in Section 1.2, and by qualified personnel.  

For additional product and services related to the use and installation of this 

OXYvisor

 product, such as technical assistance, field start-up services, 

classroom product trainings, wet-gas sample probes, heated sample lines, pressure reduction stations or other needs please contact us at Barben 

Analytical or a factory qualified local representative (see back page for complete contact information).

Section 1 - General Information & Safety

1.1 Introduction - Purpose of this Manual

1�2 About the 

OXYvisor

 Product - (Overview)

Definition of the symbols for hazardous, cautionary, precautionary and general information are found and defined here.

1.3 Definition of Symbols

W A R N I N G           

Safety warning indicates a potentially or imminently hazardous situation which, if not avoided, 
may result in personal injury or environmental contamination.

!

WARNING

!

WARNING

AVERTISSEMENT          

L’avertissement de sécurité indique une situation potentiellement dangereuse ou imminente qui, si elle 
n’est pas évitée, peut entraîner des blessures ou une contamination de l’environnement.

!

WARNING

!

WARNING

C A U T I O N 

General warning, caution, that if an operating procedure which, if not strictly observed, may 
result in damage or malfunction of the analyzer.

!

CAUTION

!

CAUTION

AT T E N T I O N 

Avertissement général, mise en garde, si une procédure d’exploitation qui, si elle n’est pas 
strictement observée, peut entraîner des dommages ou un dysfonctionnement de l’analyseur.

!

CAUTION

!

CAUTION

C A U T I O N 

Caution, possibility of electric shock may occur if an operating procedure which, if not strictly 
observed, may occur.

AT T E N T I O N 

Attention, un choc électrique peut se produire si une procédure qui, si elle n’est pas strictement 
observée, peut se produire.

IEC 60417-5019 

Protective conductor terminal. (

Borne du conducteur de protection

.)

IEC 60417-5032 

Alternating Current (AC) (

Courant alternatif

)

IEC 60417-5031 

Direct Current (DC) (

Courant continu

)

 

Useful and/or important information that should not be overlooked (

Informations utiles et / ou 

importantes à ne pas négliger

)

USEFUL INFORMATION

USEFUL INFORMATION

IEC 60417-5020 

Frame or chassis terminal. (

Terminal cadre ou châssis 

)

Section 1

General Info & Safety

Summary of Contents for BARBEN ANALYTICAL OXY visor

Page 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Page 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Page 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Page 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Page 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Page 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Page 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Page 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Page 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Page 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Page 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Page 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Page 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Page 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Page 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Page 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Page 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Page 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Page 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Page 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Page 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Page 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Page 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...