background image

17

IOM (Safety) - 

Installation, Operation and Maintenance

                      Optical Oxygen Analyzer

Section 3

Installation

Figure 13.3 - Zone 1 HazLoc - Installation Control DWG - 2P0345 (p.3)

 

3 OF 3

 

C. SHIELDS

OXYvisor HAZARDOUS LOCATION

ZONE 1 CONNECTIONS

 

UNLESS

OTHERWISE

SPECIFIED

 

ALL ± .015"

 

FRACTIONAL

.000 ± .005"

.00  ± .010"

.0   ± .015"

 

APVD BY

DRAWN BY

 

DRAWING NO

DATE

SCALE

SIZE

SHEET

PART NO.

MATERIAL/DESCRIPTION

APVD

DWN

DESCRIPTION

DATE

REV

REVISIONS

 

ANGLES ± 1/4°

 

TOLERANCES

DIMENSIONS

IN INCHES

 

ECO#

 

5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740

B. WOLF

 

01/26/2018

 

2P0345

-

OXYvisor

 

PART NO.

 

D

NTS

 

FOR

 

   2       10/22/18                    INITIAL RELEASE                  CS       BW

 

OXYvisor

OPTICAL OXYGEN ANALYZER

ZONE 1 UNIT

RTD

1

TRANSMITTER

FIBER PATCH

CONNECTION

OR WAND

RELAY CONTROLLED

DEVICES (4X)

POWER SOURCE

RS485, DIGITAL INPUTS,

ANALOG OUTPUTS

EXTERNAL POWER FOR

RELAY CONTROLLED

DEVICES

      ALL WIRING THROUGH CONDUIT ENTRY POINTS MUST BE INSTALLED AND SEALED

      IN ACCORDANCE WITH NEC, CEC, OR ALL OTHER APPLICABLE LOCAL ELECTRICAL CODES.

      DIGITAL INPUTS ARE 5VDC POWERED BY ANALYZER.

      INPUT IS SWITCH MAKE/BREAK ONLY.

      INDIVIDUAL RELAYS 24VDC ONLY:

           .05 AMP PILOT DUTY, MAXIMUM

           .40 AMP RESISTIVE LOAD, MAXIMUM

      WHEN USING THE DC POWER OPTION, POWER MUST BE FROM CLASS 2 OR LPS SOURCE.

           DC POWER OPTION:          21.6-26.5 VDC, 5W

           AC POWER OPTION:          85-264 VAC, 47-63 HZ, 6W

      SPECIFIC CONDITIONS OF USE:

    UNUSED ENTRIES SHALL BE CLOSED WITH SUITABLE CERTIFIED Ex db BLANKING ELEMENTS

      WITH A MINIMUM IP66 RATING.

    FLAMEPROOF JOINTS ARE NOT INTENDED TO BE REPAIRED.  FOR REPAIR OR REPLACEMENT

      OF ANY PART, CONTACT BARBEN ANALYTICAL.

x

x

1

ZONE 1, GROUP IIC

HAZARDOUS AREA

REQUIRES PRODUCT MODEL SELECTION:

AGENCY APPROVAL, "1."

NON-HAZARDOUS AREA

1

FIBER OPTIC

BUSHING

(Ex d BARRIER)

4

3

2

2

3

4

ENCLOSURE

WAND TO

FIBER PATCH

(IF APPLICABLE)

JUNCTION

BOX

USER SUPPLIED

WIRING AND COMPONENTS

!

3.11.2 Zone 2 - Group IIC - 2P0346

Figure 14.1 - Zone 2 HazLoc - Installation Control DWG - 2P0346 (p.1)

1 OF 3

C. SHIELDS

OXYvisor HAZARDOUS LOCATION

INSTALLATION DRAWING ZONE 2

UNLESS

OTHERWISE

SPECIFIED

ALL ± .015"

FRACTIONAL

.000 ± .005"

.00  ± .010"

.0   ± .015"

APVD BY

DRAWN BY

DRAWING NO

DATE

SCALE

SIZE

SHEET

PART NO.

MATERIAL/DESCRIPTION

APVD

DWN

DESCRIPTION

DATE

REV

REVISIONS

ANGLES ± 1/4°

TOLERANCES

DIMENSIONS

IN INCHES

ECO#

5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740

B. WOLF

01/26/2018

2P0346

-

OXYvisor

PART NO.

D

NTS

FOR

2

      ALL USED CONDUIT PORTS IN ENCLOSURE AND JUNCTION BOX MUST BE SEALED WITH

      APPROPRIATELY CERTIFIED FITTINGS AND SEALANTS PER THE APPLICABLE HAZARDOUS LOCATION.

      ALL UNUSED PORTS MUST BE PLUGGED. TO ACHIEVE IP66 RATING (MAXIMUM,)

      ALL CUSTOMER SUPPLIED FITTINGS, PLUGS, AND SEALANTS MUST BE SUITABLY RATED.

      THREADED CONDUIT PORT OPTIONS:

           M25 X 1.5 6H

           3/4" NPT FEMALE

      FRONT AND REAR ENCLOSURE DOORS MUST BE SECURED WITH THE PROVIDED

      LOCKING SCREW TO PREVENT OPENING WITHOUT THE USE OF A TOOL.

      (M4 LOCKING SCREW REQUIRES M2 HEX WRENCH)

      ENCLOSURE MANUFACTURER IS LIMATHERM S.A.:

      CONTROL DRAWING: 1-Z-L4249

      COATING:  SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER

FRONT VIEW

HAZARDOUS AREA

APPROVAL LABEL

LOCATION

REAR TERMINAL

CONDUIT PORTS

FIBER OPTIC

BUSHING PORT

O-RING SEALED

RIGHT VIEW

3

LEFT VIEW

1

2

2

5

4

ENCLOSURE

JUNCTION BOX

1

3

4

5

6

7

EXTERNAL PROTECTIVE

CONDUCTOR TERMINAL

TERMINAL 

COMPARTMENT

DOOR

7

6

MODEL NO. : XD-DR120WIN-4-N3-N3-N3-N3-M24/M5-SPU-CW

                                                           OR

                       XD-DR120WIN-4-M25-M25-M25-M25-M24/M5-SPU-CW

ATEX CERTIFICATION NO. :  FTZÚ 08 ATEX 0182U

ATEX MARKING:       II 2G Ex db IIC Gb

IEC Ex CERTIFICATION NO. : FTZÚ 09.0031U

IEC Ex MARKING: Ex db IIC Gb

Ԑx

 

JUNCTION BOX MANUFACTURER IS LIMATHERM S.A.:

           CONTROL DRAWING: 2-Z-L2453

           COATING:  SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER

      THE BOTTOM JUNCTION BOX PORT IS CONSUMED BY THE FIBER OPTIC SENSOR.

      1/2" NPT FEMALE

       FIBER OPTIC PORT/BUSHING THREADS ARE M24 X 1.5 6H/6g

       ALL OXYvisor UNITS CONTAIN A BATTERY AND WILL HAVE LIVE CIRCUITS, EVEN

       WHEN NO POWER IS APPLIED.  DECLASSIFY HAZARDOUS AREAS BEFORE OPENING.

EXTERNAL PROTECTIVE

CONDUCTOR TERMINAL

   A       06/26/19                    INITIAL RELEASE                  MW       BW

Summary of Contents for BARBEN ANALYTICAL OXY visor

Page 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Page 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Page 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Page 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Page 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Page 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Page 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Page 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Page 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Page 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Page 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Page 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Page 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Page 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Page 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Page 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Page 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Page 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Page 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Page 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Page 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Page 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Page 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Page 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Reviews: