17
IOM (Safety) -
Installation, Operation and Maintenance
Optical Oxygen Analyzer
Section 3
Installation
Figure 13.3 - Zone 1 HazLoc - Installation Control DWG - 2P0345 (p.3)
3 OF 3
C. SHIELDS
OXYvisor HAZARDOUS LOCATION
ZONE 1 CONNECTIONS
UNLESS
OTHERWISE
SPECIFIED
ALL ± .015"
FRACTIONAL
.000 ± .005"
.00 ± .010"
.0 ± .015"
APVD BY
DRAWN BY
DRAWING NO
DATE
SCALE
SIZE
SHEET
PART NO.
MATERIAL/DESCRIPTION
APVD
DWN
DESCRIPTION
DATE
REV
REVISIONS
ANGLES ± 1/4°
TOLERANCES
DIMENSIONS
IN INCHES
ECO#
5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740
B. WOLF
01/26/2018
2P0345
-
OXYvisor
PART NO.
D
NTS
FOR
2 10/22/18 INITIAL RELEASE CS BW
OXYvisor
OPTICAL OXYGEN ANALYZER
ZONE 1 UNIT
RTD
1
TRANSMITTER
FIBER PATCH
CONNECTION
OR WAND
RELAY CONTROLLED
DEVICES (4X)
POWER SOURCE
RS485, DIGITAL INPUTS,
ANALOG OUTPUTS
EXTERNAL POWER FOR
RELAY CONTROLLED
DEVICES
ALL WIRING THROUGH CONDUIT ENTRY POINTS MUST BE INSTALLED AND SEALED
IN ACCORDANCE WITH NEC, CEC, OR ALL OTHER APPLICABLE LOCAL ELECTRICAL CODES.
DIGITAL INPUTS ARE 5VDC POWERED BY ANALYZER.
INPUT IS SWITCH MAKE/BREAK ONLY.
INDIVIDUAL RELAYS 24VDC ONLY:
.05 AMP PILOT DUTY, MAXIMUM
.40 AMP RESISTIVE LOAD, MAXIMUM
WHEN USING THE DC POWER OPTION, POWER MUST BE FROM CLASS 2 OR LPS SOURCE.
DC POWER OPTION: 21.6-26.5 VDC, 5W
AC POWER OPTION: 85-264 VAC, 47-63 HZ, 6W
SPECIFIC CONDITIONS OF USE:
UNUSED ENTRIES SHALL BE CLOSED WITH SUITABLE CERTIFIED Ex db BLANKING ELEMENTS
WITH A MINIMUM IP66 RATING.
FLAMEPROOF JOINTS ARE NOT INTENDED TO BE REPAIRED. FOR REPAIR OR REPLACEMENT
OF ANY PART, CONTACT BARBEN ANALYTICAL.
x
x
1
ZONE 1, GROUP IIC
HAZARDOUS AREA
REQUIRES PRODUCT MODEL SELECTION:
AGENCY APPROVAL, "1."
NON-HAZARDOUS AREA
1
FIBER OPTIC
BUSHING
(Ex d BARRIER)
4
3
2
2
3
4
ENCLOSURE
WAND TO
FIBER PATCH
(IF APPLICABLE)
JUNCTION
BOX
USER SUPPLIED
WIRING AND COMPONENTS
!
3.11.2 Zone 2 - Group IIC - 2P0346
Figure 14.1 - Zone 2 HazLoc - Installation Control DWG - 2P0346 (p.1)
1 OF 3
C. SHIELDS
OXYvisor HAZARDOUS LOCATION
INSTALLATION DRAWING ZONE 2
UNLESS
OTHERWISE
SPECIFIED
ALL ± .015"
FRACTIONAL
.000 ± .005"
.00 ± .010"
.0 ± .015"
APVD BY
DRAWN BY
DRAWING NO
DATE
SCALE
SIZE
SHEET
PART NO.
MATERIAL/DESCRIPTION
APVD
DWN
DESCRIPTION
DATE
REV
REVISIONS
ANGLES ± 1/4°
TOLERANCES
DIMENSIONS
IN INCHES
ECO#
5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740
B. WOLF
01/26/2018
2P0346
-
OXYvisor
PART NO.
D
NTS
FOR
2
ALL USED CONDUIT PORTS IN ENCLOSURE AND JUNCTION BOX MUST BE SEALED WITH
APPROPRIATELY CERTIFIED FITTINGS AND SEALANTS PER THE APPLICABLE HAZARDOUS LOCATION.
ALL UNUSED PORTS MUST BE PLUGGED. TO ACHIEVE IP66 RATING (MAXIMUM,)
ALL CUSTOMER SUPPLIED FITTINGS, PLUGS, AND SEALANTS MUST BE SUITABLY RATED.
THREADED CONDUIT PORT OPTIONS:
M25 X 1.5 6H
3/4" NPT FEMALE
FRONT AND REAR ENCLOSURE DOORS MUST BE SECURED WITH THE PROVIDED
LOCKING SCREW TO PREVENT OPENING WITHOUT THE USE OF A TOOL.
(M4 LOCKING SCREW REQUIRES M2 HEX WRENCH)
ENCLOSURE MANUFACTURER IS LIMATHERM S.A.:
CONTROL DRAWING: 1-Z-L4249
COATING: SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER
FRONT VIEW
HAZARDOUS AREA
APPROVAL LABEL
LOCATION
REAR TERMINAL
CONDUIT PORTS
FIBER OPTIC
BUSHING PORT
O-RING SEALED
RIGHT VIEW
3
LEFT VIEW
1
2
2
5
4
ENCLOSURE
JUNCTION BOX
1
3
4
5
6
7
EXTERNAL PROTECTIVE
CONDUCTOR TERMINAL
TERMINAL
COMPARTMENT
DOOR
7
6
MODEL NO. : XD-DR120WIN-4-N3-N3-N3-N3-M24/M5-SPU-CW
OR
XD-DR120WIN-4-M25-M25-M25-M25-M24/M5-SPU-CW
ATEX CERTIFICATION NO. : FTZÚ 08 ATEX 0182U
ATEX MARKING: II 2G Ex db IIC Gb
IEC Ex CERTIFICATION NO. : FTZÚ 09.0031U
IEC Ex MARKING: Ex db IIC Gb
Ԑx
JUNCTION BOX MANUFACTURER IS LIMATHERM S.A.:
CONTROL DRAWING: 2-Z-L2453
COATING: SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER
THE BOTTOM JUNCTION BOX PORT IS CONSUMED BY THE FIBER OPTIC SENSOR.
1/2" NPT FEMALE
FIBER OPTIC PORT/BUSHING THREADS ARE M24 X 1.5 6H/6g
ALL OXYvisor UNITS CONTAIN A BATTERY AND WILL HAVE LIVE CIRCUITS, EVEN
WHEN NO POWER IS APPLIED. DECLASSIFY HAZARDOUS AREAS BEFORE OPENING.
EXTERNAL PROTECTIVE
CONDUCTOR TERMINAL
A 06/26/19 INITIAL RELEASE MW BW