5-4 | 303B Moisture Analyzer Manual
the ribbon plug from J2.
3. Remove the four mounting screws and board from its mounting.
Main AC Fuse (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Locate fuse F1 on the power supply board near the transformer.
3. Lift out F1 and test with an ohm meter. If open, use appropriate fuse
type stated in Section 5.
4. If fuse blows again, replace power supply pc board.
Battery (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Remove the wire terminals from the battery. Note that red wire is on
+ terminal.
3. Remove the two bracket nuts, move the bracket to the side, and lift
out the battery.
4. If new battery does not fix problem, check polarity. Allow 24 hours for
initial charge.
On battery replacement, use only AMETEK 303020001, nominal
12Vdc @ 5 Amp hour.
CPU/Display Board
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Remove connectors P6, P1, P3, and P5.
3. Remove the three mounting nuts and the board.
!
WARNING
Summary of Contents for 303B
Page 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Page 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Page 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Page 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Page 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Page 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Page 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Page 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Page 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Page 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Page 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Page 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Page 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...