Installation
|
3-7
Pressure / Vacuum Pump Use
If the monitor is used to sample moisture in toxic or flammable gases, the GAS OUT must
exhaust to an area deemed safe and appropriate by the local authority having jurisdiction.
Due to the SAMPLE ADJUST flow controller restriction, if your sample pressure is less
than 69 kPa (10 psi) gauge, you must increase the pressure with either a pressure (option
A in figure 3-2) or a vacuum pump (option B). An AMETEK pump (PN
271803001
) or
(271813001, 240V) can be used in either configuration. Keep in mind that some pressure
pumps will add moisture to the sample. If you plan to run your monitor with a vacuum
pump, proceed as follows:
1. Disconnect the line from the GAS IN or OUT fitting and interconnect it
with a tube union. Connect the pump to the union. Check the connec-
tion for leaks.
2. Connect the pump outlet or the GAS OUT fitting to vent in a safe area
if sampling a hazardous gas.
!
WARNING
90
to
130
Vac
Panel
180
to
260
Vac
Panel
POWER
IN
90
‐
130
Vac
47
‐
63
Hz
20W
MAX
POWER
IN
180
‐
260
Vac
47
‐
63
Hz
20W
MAX
Figure 3-1
303B - Rear Panel.
Summary of Contents for 303B
Page 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Page 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Page 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Page 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Page 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Page 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Page 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Page 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Page 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Page 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Page 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Page 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Page 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...