5-8 | 303B Moisture Analyzer Manual
In step 12, P x.xx represents your software revision number, which can be
helpful to identify software upgrades. This should be the same number
written on the cpu/display board cpu PROM.
Parameter Settings
Offset
Offset subtracts from the moisture readings.
1.
Press the POWER key (LED OFF). Hold the STANDBY key (LED
ON) and press the POWER key (LED ON) until the STANDBY, +, and –
LED’s flash.
2.
Press the + or – key to increment or decrement the display to the
desired valve.
3.
When the desired value is displayed, press OFFSET to store the
value; ProG will be displayed until the value is stored, followed by the
normal LED/display test.
If the ALARM, CELL TEST, 0-1000 (50), 0-2000 (100), or DISPLAY
LIGHT key is pressed during the routine, the new value is not stored
and normal moisture measuring begins.
Gain
Gain adjustment enables a value to be entered to compensate for non-
standard ranges or flow limitations.
1. Press the POWER key (LED OFF). Hold the ELEC TEST key (LED ON)
and press the POWER key (LED ON) until the STANDBY, +, and –
LED’s flash.
2. Press the + or – key to increment or decrement the display to the de-
sired valve.
3. When the desired value is displayed, press GAIN to store the value;
ProG will be displayed until the value is stored, followed by the nor-
mal LED/display test.
NOTE
Summary of Contents for 303B
Page 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Page 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Page 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Page 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Page 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Page 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Page 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Page 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Page 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Page 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Page 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Page 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Page 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...