Maintenance
|
5-3
Parts Replacement
Flow Meter Tubes (Figure 5-1)
1. Turn off power using front panel switch and disconnect the unit from
the power source.
2. Shut off sample flow at the source and purge the line. Shut off purge.
3. Remove the screws and lift off the cover.
4. Disconnect the “U” tube between the top fittings.
5. Remove the two screws and the plastic cover from the assembly.
6. With a hex wrench, loosen the plug over the glass tube in the assem-
bly top and remove the tube. Using a piece of tape on the glass makes
replacement easier.
7. Be sure the glass tube is centered on the top and bottom seals and the
grounding wire is around the tube before tightening the top plug.
Do not tighten the screws on the plastic cover or the cover might
break.
Flow Controller (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Loosen the two tubing ferrule nuts at the flow controller and discon-
nect the lines.
3. Remove the mounting nuts and lift out the flow controller and brack-
et.
4. Remove the screw and remove the flow controller from the bracket.
Power Supply PC Board (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Loosen the screws and remove the power-in wires from J1; remove
NOTE
Summary of Contents for 303B
Page 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Page 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Page 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Page 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Page 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Page 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Page 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Page 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Page 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Page 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Page 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Page 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Page 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...