5-10 | 303B Moisture Analyzer Manual
Table 5-1. Troubleshooting Chart
Symptom
Possible
Cause
Corrective
Action
No display
Power cord not plugged in
Plug in power cord
Blown Fuse
Replace fuse (F1)
Faulty power supply board
Replace power supply
board.
Display reads over range Temporarily saturated cell
If normal reading is
not
obtained
in
1
hour
continue drydown.
Saturated cell
Shut off sample flow at the
source and allow cell to dry.
Keep
303B
operating
electrically.
Display reading erratic
Shorted cell
Replace the cell
(noisy)
Varying sample flow
Replace the flow regulator
Intermittent short
Replace the cell
Faulty power supply board Replace power supply board
Display reading low on
Low gas flow rate
Check and adjust flow rate
gas with a known moisture
content
Clogged filter or dirty
Replace filter. Replace flow
sample flow valve
regulator
Contaminated or dirty cell, Remove, flush and recoat cell
or
partially
inactive
phosphorous
pentoxide
Display reading high High gas flow rate
Check and adjust flow rate
on gas with a known
moisture content
Leak in sample system
Do a leak test
Dirty flow meter
Replace flow regulator
Shorted or partially
Remove, flush and recoat cell
Summary of Contents for 303B
Page 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Page 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Page 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Page 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Page 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Page 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Page 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Page 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Page 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Page 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Page 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Page 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Page 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Page 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Page 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...