MEASure:CURRent:AMPLitude:MAX?
This query returns the peak value of the output AC current being sourced at the output
terminals. Note that the output relay must be closed to obtain current flow. The peak current
measurement circuit uses a sample and hold method and latches the highest peak current
value found since the last peak measurement reset command.
Query Syntax
MEASure:CURRent:AMPLitude:MAX?
Parameters
None
Examples
MEAS:CURR:AMPL:MAX?
Returned Parameters
<NR2>
Related Commands
OUTP 0 1
MEAS:CURR?
MEASure:CURRent:AMPLitdue:RESet
This command resets the peak current measurement sample and hold circuit to zero.
Query Syntax
MEASure:CURRent:AMPLitude:RESet
Parameters
None
Examples
MEAS:CURR:AMPL:RES
Related Commands
MEAS:CURR:AMPL:MAX?
MEASure:CURRent:CREStfactor?
This query returns the crest factor of the output AC current being sourced at the output
terminals. Crest factor is the ratio of peak value to rms value. Note that the output relay
must be closed to obtain current flow.
Query Syntax
MEASure:CURRent:CREStfactor?
Parameters
None
Examples
MEAS:CURR:CRES?
Returned Parameters
<NR2>
Related Commands
MEAS:CURR? MEAS:CURR:AMPL:MAX?
MEASure:CURRent:RANGe
This commands selects the measurement range to use for current, peak current and power
measurements. Available choices are high range (0), low range (1) or auto ranging (2). See
measurement specifications (section 2.3) for actual range values.
Query Syntax
MEASure:CURRent:RANGe
Parameters
0 | 1 | 2
Examples
MEAS:CURR:RANG?
Returned Parameters
<NR2>
Related Commands
MEAS:CURR? MEAS:CURR:AMPL:MAX?