[SOURce:]FREQuency
This command sets the frequency of the output sinewave.
Command Syntax
[SOURce:]FREQuency <NRf>
Parameters
range specified by the LIM:FREQ:LOW and LIM:FREQ:HIGH
commands
Unit
Hz (Hertz)
*RST Value
defined by PON
Examples
FREQ 50
Query Syntax
FREQuency?
Returned Parameters
<NR2>
Related Commands
LIM:FREQ;LOW? LIM:FREQ:HIGH?
[SOURce:]VOLTage
This command programs the AC rms output voltage level of the power source. The
maximum voltage value allowed is determined by the selected voltage range. For 2003RP-
AV models configured for single voltage range, only one range is available.
Command Syntax
[SOURce:]VOLTage[:LEVel] <NRf>
Parameters
0 to maximum voltage range specified by the VRAN command
Unit
V (rms voltage)
*RST Value
0 volt
Examples
VOLT 250
VOLT:LEV 25
Query Syntax
[SOURce:]VOLTage[:LEVel]?
Returned Parameters
<NR2>
Related Commands
VOLT:RANGe