6.4.3
RO HEAT DISINFECT button
Takes you to the RO HEAT DISINFECT screen shown below.
Here specific functions for the RO heat disinfect can be set.
TANK TEMP determines the maximum desired tank temperature during a heat
disinfection cycle (max 185°F or 85°C).
TANK RETURN TEMP determines the actual disinfection temperature, maximum
185°F (85°C). Tank return temp should always be set 2°F below tank temp.
HOLD PERIOD is the time water is circulated around the system once it has
reached the TANK RETURN TEMP i.e. the disinfection temperature.
TIME REMAINING shows time remaining to complete the hold period.
RO HEAT DISINFECT TIME SCHEDULE takes you to the screen shown below and
enables automatic operating times to be set. System operation is activated when
TIMER is selected in the OPERATION MENU screen.
Figure 41
Figure 42