36
98-0157 Rev K
If both the OFF button and the function button remain illuminated, the function
selected is completing shut down processes, when these are complete the
function button will turn gray. It will not be possible to select other functions
until only the red OFF button is illuminated red.
If a manual RO HEAT DISINFECT is selected it can be aborted by
pressing the OFF button.
MAIN MENU button
Switches screen to Main Menu
STATUS MONITOR button
Switches screen to status monitor (see Figure 26)
SETTING MENU button
Switches screen to Settings Menu (see Figure 32)
CALIBRATION MENU button
Only accessible by AmeriWater technicians.
6.1.2
Operation Periods
MediQA Operation Periods
Period
Explanation
Standby
The MediQA can be set to “POWER ON STANDBY” for occasions when there
are long periods (typically days during shut down periods) without use. In
standby mode the MediQA performs 10 minutes of operation in every 2
hours. The duration and interval are user settable. This enables water to be
circulated through the MediQA and distribution loop, in order to maintain
water quality.
Supply
Supply is when the RO begins providing water to the loop for use with the
Dialysis machines. This is the process the RO will be in for a majority of its
usage. Can be operated in continuous or a set schedule.
Flush
Before the MediQA provides water to the loop for dialysis use, the system
goes through a flush. This isolates the water from the loop and dumps all of
the water down the drain for a preset time.
Rinse
After the flush cycle is completed, the RO will go into a rinse. During rinse the
unit continues to run the pumps at a high pressure. Water is flushed down
the drain similar to the flush of the membranes.
Refresh
Loop has exceeded the temperature set point and is refreshing the water to
lower the temperature.
RO Heat Disinfect
The MediQA is capable of a heat disinfection of its plumbing and membranes.
During the heat disinfection process the RO circulates water within itself at
185˚F.
Temp Hold
Once the temperature is reached in the RO Heat Disinfect period, the water is
circulated for a set period. Hold time is user settable, see Section 13.1 for
default.
Summary of Contents for MediQA MDP Series
Page 70: ...69 98 0157 Rev K 10 FLOW DIAGRAMS SPH UNITS ...
Page 71: ...70 98 0157 Rev K DPH UNITS ...
Page 72: ...71 98 0157 Rev K 11 WIRING SCHEMATIC SPH UNITS ...
Page 73: ...72 98 0157 Rev K ...
Page 74: ...73 98 0157 Rev K ...
Page 75: ...74 98 0157 Rev K ...
Page 76: ...75 98 0157 Rev K ...
Page 77: ...76 98 0157 Rev K ...
Page 78: ...77 98 0157 Rev K ...
Page 79: ...78 98 0157 Rev K ...
Page 80: ...79 98 0157 Rev K DPH Units ...
Page 81: ...80 98 0157 Rev K ...
Page 82: ...81 98 0157 Rev K ...
Page 83: ...82 98 0157 Rev K ...
Page 84: ...83 98 0157 Rev K ...
Page 85: ...84 98 0157 Rev K ...
Page 86: ...85 98 0157 Rev K ...
Page 87: ...86 98 0157 Rev K ...