43
98-0157 Rev K
6.4.1
RO SUPPLY button
Screen shown below for DPH.
Screen shown below for SPH.
AUTO RINSE MIN defines the duration of initial rinse. If unit has not rinsed to
quality set-point, system will rinse for an additional minute until max time has
been reached. If unit has not reached the quality after max time elapsed an
alarm is logged.
AUTO FLUSH defines the duration of flush.
TARGET RECOVERY RO1 sets target water recovery of stage 1 RO modules.
TARGET RECOVERY RO2 sets target water recovery of stage 2 RO modules.
SUPPLY HEATING defines whether the tank heater is to be used to warm the
incoming water supply. When enabled, the heater is turned on at 48°F (9 °C)
and off at 52°F (11°C).
Figure 36
Figure 37