background image

EN

16

Insertion of the Ambu® aScope™

The Ambu® aScope™ can be inserted through the mouth or nose. When the Ambu® aScope™ is inserted, slightly advance it with the distal tip in a neutral 
position. View the image continuously on the Ambu® aScope™ Monitor when passing the distal end of the Ambu® aScope™ from the mouth/nose to the 
larynx and from the larynx to the carina. It is important to recognise the anatomical structures and avoid damage to the mucosal wall.

 

 lf resistance is encountered when inserting the Ambu® aScope™, do not attempt to use force. 

 

 When the Ambu® aScope™ is inserted through the mouth it is recommended to place a mouth piece to protect the Ambu® aScope™ 
from being damaged.

If the distal tip of Ambu® aScope™ is contaminated it can be cleaned with a piece of sterile gauze or a hospital disinfection wipe. Continue this procedure 
until a satisfactory image is obtained.

12.1 Injection and control of the Luer channel

The Ambu® aScope™ has a Luer channel where it is possible to inject topical anaesthesia.
The Luer is compatible with all syringes with ISO connection. It is recommended that the Luer channel is closed when it is not in use. Insert a syringe of 
topical local anaesthetic into the Luer channel and press the plunger. To ensure that all the local anaesthetic has left the channel, flush  
the channel with 2ml air.

12.2 Removal procedure

 Slowly withdraw the Ambu® aScope™ while observing the image on the Ambu® aScope™ Monitor. 

 

  The distal tip must be in a neutral and non-deflected position. Otherwise there is a risk that the Ambu® aScope™ can be damaged and/or 
the patient may be injured.

  

 After the Ambu® aScope™ has been used check for damage or missing parts before it is placed in a waste container.

 

 Disconnect the Ambu® aScope™ from the Ambu® aScope™ Monitor and dispose of the Ambu® aScope™ in accordance with local 
guidelines for collection of infected medical devices with electronic components.

If the Ambu® aScope™ is used more than once on the same patient during the same procedure, turn the Ambu® aScope™ off in between sessions and 
place it on a sterile surface. 

 

 Be aware of the total operating time limit of 30 minutes during a period of 8 hours from first switching on.

12.3 Post-check guide

 

 A visual check as described below should be carried out before finalising the procedure and disposing of the Ambu® aScope™. If any test 
fails, take corrective action in order to reduce trauma to the patient.

Visual test – Ambu® aScope™

1.  Are there any missing parts on the bending section, lens, or insertion cord? If yes, then take corrective action to locate the missing part.
2.   Is there any evidence of damage on the bending section, lens, or insertion cord? If yes, then examine the integrity of the product and conclude  

if there are any missing parts.

3.   Are there cuts, holes, sagging, swelling or other irregularities on the bending section, lens, or insertion cord? If yes, then examine the product to  

conclude if there are any missing parts.

In case of corrective actions needed (step 1 to 3) act according to local hospital procedures. The elements of the insertion cord are radiopaque

12.4 Cleaning of the Ambu® aScope™ Monitor:

The Ambu® aScope™ Monitor must be cleaned and disinfected according to the instructions before first use.

1.    Prepare a cleaning solution using a standard enzymatic detergent (Enzol or equivalent) prepared per manufactures recommendations. 

Recommended detergent: enzymatic, mild pH: 7-9, low foaming. Contact Ambu A/S for further information on recommended detergents.

2.   Soak a sterile gauze in the enzymatic solution and then wring out to ensure that the gauze is not dripping.
3.    Thoroughly clean the buttons, screen and external casing of the monitor with the damp gauze. Avoid getting the device wet to prevent damaging 

internal electronic components.

4.   Using a sterile soft bristled brush that has been dipped in the enzymatic solution, brush the buttons until all evidence of soil is removed.
5.   Wait for 10 minutes to (or the time recommended by the manufacturer of the detergent) allow the enzymes to activate.
6.   Rinse the device using sterile gauze that has been dampened with RO/DI water. Ensure all evidence of the detergent is removed.
7.   Repeat steps 1 to 6

Summary of Contents for aScope

Page 1: ...ns for use Ambu aScope For Single Patient use only Sterile For use by trained clinicians physicians only For in hospital use Ambu aScope Monitor For use by trained clinicians physicians only For in ho...

Page 2: ...800 ambude www ambu de UK Ambu Ltd 8 Burrel Road St Ives Cambridgeshire PE27 3LE Tel 44 1480 498 403 Fax 44 1480 498 405 www ambu com Italy Ambu S R L Via Paracelso 18 Centro Direzionale Colleoni 2004...

Page 3: ...to an external monitor video recorder 15 10 Clinical Application 15 11 Preparation for Use 15 12 Use of the Ambu aScope system 16 12 1 Injection and control of the Luer channel 17 12 2 Removal proced...

Page 4: ...loose parts or the bending section does not work as intended Do not touch the distal tip of the insertion cord or allow it to strike other objects The lens surface of the distal tip is fragile and vis...

Page 5: ...conventional cleaning and sterilization procedures 2 Intended Use The Ambu aScope is intended for use as an aid in the placement of an Endotracheal Tube ETT directly or through an intubating laryngea...

Page 6: ...e Ambu aScope Working length of the Ambu aScope insertion cord Maximum width of inserted portion Min ETT ID 6 mm Minimum ET tube inner diameter 6 0mm see section 12 Field of view Operating time of the...

Page 7: ...and electrical waste WEEE Only applicable for the Monitor Li ion Battery type Lithium ion Only applicable for the battery inside the Ambu aScope Monitor Re chargeable battery Only applicable for the...

Page 8: ...duct specifications Ambu aScope Optical System Field of View 80 Direction of View 0 forward viewing Depth of Field 2mm 50 mm 0 1 1 968 inch Illumination method LED LTW C282DS5 SE Insertion cord Distal...

Page 9: ...al standard NTSC Video out connection 3 5mm jack socket Video out cable 3 5mm jack plug to RCA jack plug Storage Temperature 10 40 C 50 104 F Relative humidity 10 90 Mounting interface Mounting interf...

Page 10: ...r battery icon on the screen changes from green fully charged battery via orange half battery capacity to red low bat tery capacity within 30 minutes it is recommended that the Ambu aScope Monitor is...

Page 11: ...t Function Material 1 Monitor frame PC Polycarbonate ABS Acrylonitrile Butadiene 2 Monitor screen Shows the picture from the camera Glass 3 Bracket Secures the monitor to e g an IV pole POM GF25 Polyo...

Page 12: ...sportation purposes Cardboard Table 5 Functions and materials 8 2 1 Monitor placement on the bracket The Ambu aScope Monitor can be placed on a flat surface But to secure the Ambu aScope Monitor it is...

Page 13: ...alf time left orange Time is expiring Less than 5 minutes left red Plug is connected and the monitor battery is above half charged green Plug is connected and the monitor battery is half charged orang...

Page 14: ...device should not be used A physician who is familiar with this device can perform the preparation for use in a time range of 1 4 minutes The tests should be conducted in a manner consistent with acce...

Page 15: ...tructions do not explain or discuss clinical intubation procedures They describe only the basic operation and precautions related to the operation of this Ambu aScope system Before initial use of the...

Page 16: ...rocedure turn the Ambu aScope off in between sessions and place it on a sterile surface Be aware of the total operating time limit of 30 minutes during a period of 8 hours from first switching on 12 3...

Page 17: ...d collection schemes for disposal of batteries In order to remove the battery use a Torx screwdriver to remove the battery lid 13 Trouble shooting guide If problems occur with the system please use th...

Reviews: