en

c) 

Remove the 2 screws holding each VDI board using a Phillips screwdriver

d) 

Pull out the VDI boards from the displaying unit and dispose them according to 
local guidelines for disposal of electronic waste

13

Summary of Contents for aBox 2

Page 1: ...For use by qualified biomedical technicians only Ambu aBox 2 SPARE PART REPLACEMENT MANUAL...

Page 2: ...Ambu is a registered trademark and aScope and aBox are trademarks of Ambu A S 2797 2...

Page 3: ...utions 4 1 3 Data Security 5 1 4 System Description 6 1 5 Spare Parts 7 1 6 Spare Part Replacement Process 8 2 How To Clean and Disinfect the Displaying Unit 9 3 Preparing for Spare Part Replacement 1...

Page 4: ...start replacing any spare part of the displaying unit unless you have read through this document and are qualified in the following areas Knowledge experience and acquaintance with electronic repair a...

Page 5: ...parts supplied by Ambu Do not modify the spare parts Cleaning and disinfection wipes shall be moist but not dripping to ensure no damage to internal electronics of the displaying unit If using wipes c...

Page 6: ...System Description The displaying unit is intended to display live imaging data from Ambu visualization devices 7 8 19 18 17 16 14 15 13 12 10 11 1 2 3 4 5 6 9 1 3 2 4 Wi Fi Figure 1 System illustrati...

Page 7: ...otated 10 Video output ports 2 x DVI D Enable connection to external monitor or external medical imaging recorder 11 Video output ports 2 x 3G SDI 12 Wi Fi antenna connector Enables connection of Wi F...

Page 8: ...these Instructions for Use carefully before performing any replacements on the Ambu aBox 2 The technician and personnel being responsible for the replacement process should fulfil the requirements out...

Page 9: ...following use Excessive soil should be removed on accessible areas of the device excluding electrical ports Limitations The displaying unit is not compatible with ultrasonic or automatic cleaners and...

Page 10: ...ide of rubber covers screen external casing of the monitor with the moist gauze Avoid getting the displaying unit wet to prevent damaging internal electronic components 4 Wait for 10 minutes or the ti...

Page 11: ...ys use ESD protection when disassembling replacing a spare part and reassembling the displaying unit To avoid malfunction of the equipment only use spare parts supplied by Ambu Do not modify the spare...

Page 12: ...at the original positions a Detach the front cover from the displaying unit Locate the groove on one of the sides of the front cover where a snap feature is located on the inside Press inwards with t...

Page 13: ...Remove the 2 screws holding each VDI board using a Phillips screwdriver d Pull out the VDI boards from the displaying unit and dispose them according to local guidelines for disposal of electronic was...

Page 14: ...h the supplied screws using the Phillips screwdriver into their original positions g Attach the new front cover supplied with the replacement kit Start by attaching from one end Ensure that the snap f...

Page 15: ...inspection with a functioning scope Connect the displaying unit to a suitable power outlet via the power cable II Connect a functioning visualization device to the displaying unit by plugging it into...

Page 16: ...495407500 V02 2022 04 Ambu A S Baltorpbakken 13 2750 Ballerup Denmark T 45 72 25 20 00 ambu com...