1. Important Information
This is a Spare Part Replacement Manual for Ambu® aBox™ 2. It is applicable to aBox 2 with
item number 505001000 only.
The following terms are used in this Spare Part Replacement Manual:
•
“Displaying unit” refers to Ambu® aBox™ 2.
•
“Visualization device” refers to compatible Ambu endoscopes and other compatible
visualization devices that can be connected to and used with the displaying unit.
This manual may be updated without further notice. Contact your local Ambu representative
or download the most recent version from ambu.com found om the download section on the
product page.
Please observe that the warranty will be void, if parts of the displaying unit has been changed
within the warranty period without the written consent from Ambu.
1.1. Qualification Requirements for Personnel Performing Spare Part
Replacement
Do not start replacing any spare part of the displaying unit, unless you have read
through this document and are qualified in the following areas:
•
Knowledge, experience and acquaintance with electronic repair and safety testing
according to IEC 60601-1
(Medical electrical equipment, Part 1: General requirements for basic safety and essential
performance) and
IEC 62353 (Medical electrical equipment – Recurrent test and test after repair of medical
equipment).
•
Knowledge of and experience with local regulations.
•
Having read the Instruction for Use for the displaying unit (find the Instruction for Use in
the sales package or download the current version from ambu.com).
•
Knowledge of the environment where the displaying unit is installed and used.
•
1.2. Warnings and Cautions
WARNINGS
•
To avoid risk of electric shock, always disconnect the power cable from the
displaying unit before performing spare part replacement.
•
To avoid contamination, always clean and disinfect the displaying unit before and
after spare part replacement and wear gloves during handling.