en
CAUTIONS
•
To avoid harming the electrical components, always use ESD protection when
disassembling, replacing a spare part, and reassembling the displaying unit.
•
To ensure the displaying unit functions optimally and to avoid compromising
electrical safety, always perform the tests specified in section 5 after it has been
reassembled. If any part of the tests fails after spare part replacement, do not use the
displaying unit.
•
To avoid malfunction of the equipment, only use spare parts supplied by Ambu. Do
not modify the spare parts.
•
Cleaning and disinfection wipes shall be moist, but not dripping to ensure no
damage to internal electronics of the displaying unit.
•
If using wipes containing hypochlorite or citric acid for cleaning, ensure that all
residue is completely removed. Wipes containing hypochlorite or citric acid may
affect the screen's antireflective coatings over time. You should limit the use of wipes
containing hypochlorite or citric acid to required cases only.
1.3. Data Security
To avoid possible loss of data, it is recommended to export important files before performing
any spare part replacement (please refer to the
Instruction for Use
).
5