P1500W Series Air Dryers User’s Guide
Page 26 of 101
P011787 – Rev. R
Identifying Dryer Alarms
1.1.1
High Outlet Pressure Alarm -
Occurs when the Outlet Pressure
(
OUTP
) rises above the alarm set
point for more than one (1)
minute. (Default setting is 12.0
PSI for W models, 7.50 PSI for WLP models & 35.0 PSI for WHP models)
See section 13.5 for troubleshooting information.
8.2.6
Low Outlet Pressure Alarm –
Occurs when the Outlet Pressure
(
OUTP
) drops below the alarm
set point for more than one (1)
minute. (Default setting is 6.5
PSI for W models, 0.30 PSI for WLP models & 25.0 PSI for WHP models)
See section 13.7 for troubleshooting information.
8.2.7
High Flow Rate Alarm –
Occurs when the Flow Rate
(
FLOW
) rises above the
alarm set point for more than
one (1) minute. (Default
setting is 500 SCFD)
8.2.8
High Humidity Alarm –
Occurs when the Humidity
level rises above the alarm set
point for more than one (1)
minute.