Measurement and Report
2-45
❚
Elasto Strain Ratio
NOTE
Elasto Strain Ratio is only available to use on freezing in Elastography mode.
1
1
Select
Elasto
on the touch screen to enter Elastography mode.
Live Dual mode is displayed.
2
2
Perform the scan. Proper manual compression/decompression is displayed by the
colored strain map.
3
3
Press the
[Freeze]
key to freeze the image.
4
4
Press the
[Caliper]
key.
5
5
Select
Elasto Strain Ratio
on the touch screen.
6
6
Use
[Trackball]
to move the marker to the start point on the Elastography image, and
press the
[Set]
key.
The start point is fixed, and the end point appears.
7
7
Use
[Trackball]
to move the marker to the end point, and press the
[Set]
key.
The target strain is shown on the
Result
window.
8
8
The copied caliper appears on the Elastography image area.
9
9
Use
[Trackball]
to move the marker of the copied caliper to the reference strain point
on the Elastography image, and press the
[Set]
key.
The reference strain is shown on the
Result
window.
10
10
The measured value is fixed. The strain ratio is shown on the
Result
window.
❚
Volume Flow Area
1
1
Select
Volume Flow Area
on the touch screen.
2
2
Use
[Trackball]
to move the marker to the start point.
3
3
Press the
[Set]
key to fix the point. The end point, overlapping the start point,
appears.
4
4
Use
[Trackball]
to move the marker to the end point, and press the
[Set]
key.
5
5
The measured value is fixed, and the volume flow area is shown on the
Result
window.