![Alpinion Medical Systems X-CUBE 70 Quick Manual Download Page 85](http://html1.mh-extra.com/html/alpinion-medical-systems/x-cube-70/x-cube-70_quick-manual_2908692085.webp)
Measurement and Report
2-39
❚
Angle
Angle 3 Points
1
1
Select
Angle
on the touch screen. The start point appears.
2
2
Use
[Trackball]
to move the marker to the start point, and press the
[Set]
key.
The first point is fixed, and the second point appears.
3
3
Use
[Trackball]
to move the marker to the second point.
4
4
Press the
[Set]
key. The third point appears.
5
5
Use
[Trackball]
to move the third point to the end point of the angle measurement.
6
6
Press the
[Set]
key. The angle between the two lines appears.
7
7
The angle (
Angle
) is shown on the
Result
window.
Angle 3 Lines
1
1
Select
Angle
on the touch screen. The start point appears.
2
2
Use
[Trackball]
to move the marker to the start point, and press the
[Set]
key.
The first point is fixed, and the second arrow-shaped point appears.
3
3
Use
[Trackball]
to move the marker to the second point.
4
4
Draw the baseline, and press the
[Set]
key to fix the baseline.
The third point appears.
5
5
Draw the remaining two straight lines to cross the baseline as the same way.
The α angle and β angle appear.
6
6
Press the
[Set]
key at the end point of the third straight line to fix the measured
value. The two angles (
Alpha
,
Beta
) are shown on the
Result
window.