Measurement and Report
2-41
❚
Volume
3 Distance volume
1
1
Select
Volume
on the touch screen. The start point appears.
2
2
Use
[Trackball]
to move the marker to the start point, and press the
[Set]
key.
The first point is fixed, and the second point appears.
3
3
Use
[Trackball]
to move the marker to the second point, and press the
[Set]
key
again.
The first distance is measured.
4
4
Measure the second and third distances in the same manner.
When the three distances are measured, the volume is calculated.
5
5
The measured value is fixed. The three distances (
D1
,
D2
,
D3
) and the volume (
Vol
)
are shown on the
Result
window.
Ellipse
1
1
Select
Volume
on the touch screen. The start point appears.
2
2
Use
[Trackball]
to move the marker to the start point, and press the
[Set]
key.
The first point is fixed, and the second point appears.
3
3
Use
[Trackball]
to move the marker to the second point and press the
[Set]
key. The
ellipse appears.
4
4
Use
[Trackball]
to adjust the height of the ellipse, and press the
[Set]
key again.
–
To increase the height, move
[Trackball]
up and right.
–
To decrease the height, move
[Trackball]
down and left.
5
5
The measured value is fixed. The two distances (
D1
,
D2
) and the volume (
Vol
) are
shown on the
Result
window.