1. The increased conversion rate only takes effect after a measur-
ing operation has been initiated; before this, the device remains
set at 10 measuring operations per second.
2. During measuring operations at a higher measuring rate, the
ALMEMO connectors are no longer monitored; i.e. the connec-
tor configuration can only be modified when the measuring de-
vice has stopped.
3. The printout of alarm values is not provided.
4. Analog output is no longer provided.
5. At rates above 10 measuring operations per second, mains
hum suppression is no longer provided; and as a result, accu-
racy may be adversely affected by interference over the connec-
tion lines; (wherever possible used twisted wires!).
7.5 Time and Date of Start, Time and Date of End
A sequence of measurements can, at certain points in time, be automatically
started and stopped. For this purpose the start time and date and the end time
and date can be programmed. If a date has not been specified the
measurement will be performed every day in the defined period. The current
time must be programmed beforehand.
Function ´Start Time´:
StartTime: 07:00:00
Function ´End Time´:
EndTime: --------
Input (see 5.5) in format:
hh:mm:ss
Function ´Start Date´:
StartDate:01.05.00
Function ´End Date´:
EndDate: --------
Input (see 5.5) in format:
dd:mm:yy
To clear values, select function and then use:
<
CLR
>
If the start time of a measurement has been
programmed, the status line will include the symbol:
´
´
If the end time of a measurement has been
programmed, the status line will include the symbol:
´
´
ALMEMO
®
5990-2
45
Times and Cycles