6.4.2 Averaging Mode
A detailed description of averaging over measuring point scans is given in the
manual section 6.7.4. The type of averaging is determined through the function
´Averaging Mode´. If a sensor with an ALMEMO
®
connector is connected the
following modes can be set (see 5.5):
Function no averaging:
Averag.Mode:-----
Averaging over all measuring point scans:
CONT
M
¯
Averaging over all scans of a print cycle:
CYCL
Average value of all measurements from start to stop:
STSTP
Lightening up for control purposes if averaging is in progress:
M
¯
6.4.3 Averaging over Manual Single Measurements
Single manual meas. point scans Ex are performed for an averaging of
isolated measurements at certain points or times (e.g. net measurements
according to VDI/VDE 2640, see manual 3.5.5). At all measuring points where
measured values must be averaged the averaging must be switched on by
using the averaging mode ´CONT´, a measurement in progress must be
stopped.
E1
E 2
E3
E4
CLR
M=( E )/N
Σ
i
i
MANU
MANU
MANU
MANU
1. Set the averaging mode:
Averag.Mode:
CONT
For smoothing a measured value
select attenuation, if required:
Attenuate: 20
In addition, switch the continuous measurement off:
ConvRate: 10 Cont: -
2.
Clear the average value by selecting it and using:
<
CLR
>
Function ´Average Value´ displays:
AverageVal: -----ms
M
Function ´Count´ displays:
Count: 00000
3. Manual scans of measured values:
<
MANU
>
Function ´Average Value´ displays:
AverageVal: 12.34ms
M
Function ´Count´ displays:
Count: 00001
4. Repeat step 3 for each measuring point.
5. Output of all function values of the menu by using: <
>
32
ALMEMO
®
5990-2