are entered as dotted lines. In addition, another fast display mode is available
with indicating and overwriting all measured values at the conversion rate
from left to right.
The functions ´Analog Start´ and ´Analog End´ of the menu ´Special
Functions´ can be used to set the display range of the y axis (see 9.10.4). They
can be directly called up using the key ´SCALE´.
Displaying a measured value as line chart:
Enter the print cycle in the menu ´Cycles-Times´.
PrintCycle: 00:00:05
Time axis 120 x 5s = 10min:
00:10
Scaling of the y axis in the menu ´Line Chart´:
<
SCALE
>
In the menu ´Special Functions´ set the analog
Analog-Start: 20.0%H
start and analog end of the measuring channel.
Analog-End: 80.0%H
Returning to the menu ´
Liniengrafik
´ :
ESC
,
ESC
In menu ´Line Chart´ select the meas. channel:
<
M
▲
>
Starting the measurement:
<
START
>
´
´
Stopping the measurement:
<
STOP
>
´
´
Changing to fast display using the key:
<
TIME
>
Check the conv. rate in the menu ´Cycles-Times´.
ConvRate: 10 Cont: -
Time axis 120 x 0.1s= 012s
012s
Channel switching is blocked during the measurement!
At each start and at each channel switching the line chart will be
cleared!
6.2.5 Continuous Output/Storage of all Measuring Points
The highest recording speed can be achieved using a continuous measuring
point scan (see manual 6.5.1.3), by activating the switches ´Memory ON´ or
´Output ON´ in the function ´Conversion Rate´ (see 7.4) of the menu
´Times-Cycles´, logging all measured values that have been acquired at the
conversion rate.
6.2.6 Memory Space, Memory Output, Clearing the Memory
The function ´MemoryFree´ allows, during recording of measuring data, to
continuously monitor the available memory space given in kilobytes. By
selecting this function, two soft keys are available for a direct output and
clearing of the memory. The output format corresponds to the setting in the
print cycle (see 6.2.3 and 7.2).
Function ´MemoryFree´, e.g.:
MemoryFree:
kB
0211.0
Memory data output:
<
>
Clear memory:
PROG,
<
CMEM
>
ALMEMO
®
5990-2
27
Measuring using the measurement menus
Summary of Contents for ALMEMO V5 5990-2
Page 76: ...76 ALMEMO 5990 2 Appendix...
Page 77: ...ALMEMO 5990 2 77...