22
Technical Data / Maintenance
Measuring Range
0,0...500,0
N
Terminal adapter width
0,5 | 0,8 | 1,0 | 1,3 | 1,5 | 2,0 | 2,5 | 3,0 | 3,5 | 4,0 | 5,0 | 6,0 mm
Clamping device
0,1 ... 6,0 mm (continuously variable)
Dia. range IEC 60352-2
cross section 0,05 ... 10 mm² (AWG8...30)
Wire Diameter
Dia. range SAE AS7928 II
AWG 12 ... 28
Resolution
0,1 N
@ 23°C (F.S.)
+/- 0,5% (+/- ½ digit)
Tk (absolute)
automatic readjustment while starting
Accuracy
Tk (relative)
+/- 0,02% (°K)
Standard
displays actual value in N | kgf | lbf
Operation Mode
Peak
displays peak value in N | kgf | lbf
Overload
max. admissible
200 % (F.S.) | alarm at 120% (F.S.)
Type
LCD, 4-digit, 12mm high
Update time (Standard)
1000 msec | 500 msec | 333 msec | 200 msec
100 msec | 50 msec (selectable)
Display
Update time (Peak)
1 msec
Memory
peak value (Peak)
Power Supply
Type
internal NiCd-battery (up to 12h)
AC adapter-/charger 100…240VDC (50...60Hz)
Interface
RS232C
Baud rate selectable 2,4 kB | 4,8 kB | 9,6 kB | 19,2 kB
Operation
0°... 40° C
Temperature
Range
Storage
-20°... 60° C (rF < 80%)
Protection Code
IP 40
Weight
approx. 14 kg
Dimensions
LxWxH (without lever)
350 x 160 x 75 mm
Housing Material
anodised aluminium, surface tempered steel,
stainless steel V2A
This instrument is maintenance-free under proper use. Torn parts of wire must not fall into the spaces around
the connection adapter; should this occur, blow them off or remove them using the vacuum cleaner.
Summary of Contents for FMT-100WT
Page 2: ...2 ...
Page 3: ...3 Deutsch 4 English 14 Francais 24 Espanol 34 Italiano 44 ...
Page 55: ...55 ...
Page 56: ...56 ...