Operating Procedures
Switching on the instrument, selecting measuring unit and operational mode
Switch on the instrument by pressing the ON/OFF-button. After a power-up self test and calibration (approx. 20
sec.), the display will indicate the last selected measuring unit, function and the value 0,00.
Select the desired operation mode (indication of the current measurement value or the peak value) and the
measuring unit by pressing PEAK- und MODE-buttons. The current settings will be indicated on the display.
When you choose the PEAK mode, the peak value will be always indicated on the display (drag indicator
function).
Reset
It is necessary to zero-set the instrument before starting each tractive force
measurement. Press the RESET-button for this purpose. The RESET-button
can also be used for deleting the peak value updated in the memory.
Measuring
Select the smallest suitable slot for the test sample diameter.
Place the cable connection into the adapter, so that the sleeve of
the cable connector stays securely inside the ring and cannot be
pulled out through the slot.
The excentric with the hand lever should be at the home position
fully open.
Insert the free cable end with slight tension into the excentric and
pull the hand lever clockwise. The nut in the rotative center will
help you to find the right position for the cable.
N
0.0