Alltion (Guangxi) Instrument Co., Ltd.
AM-2000 User’s Manual | Cleaning and Maintenance of the Instrument
41
Cleaning and Maintenance of the Instrument
If possible, the equipment and accessories shall be cleaned immediately after use.
Contaminants cannot be dried on the object as this will make it more difficult to clean and
disinfect.
Preventive maintenance and inspection
Operator of preventive inspection, maintenance and calibration: User
Maintenance cycle: 3 months.
Parts for preventive inspection and maintenance: eyepiece and Objective
Clean the optical surface
The optical component has a multi-layer laminating (e.g., eyepiece, objective) to ensure
optimum image quality, the contaminants attached to the surface of the optical component will
reduce the image quality. The internal optical equipment shall be protected from dust, the
equipment shall not be stored without objective lens, binocular and eyepiece. After use, cover
the dust cover on the surface of the system to avoid dust on the equipment. When the optical
components and accessories are not in use, they are always kept in the dust-free box.
Do not use any chemical cleaning agent, corrosive solvent or detergent with scratching effect,
which will damage the surface of the optical device.
The stain (blood stains, etc.) on the surface of the optical part shall be removed with distilled
water having added an appropriate amount of washing liquid. The surface can only be
wiped with humid cloth, and the surface of the component must not be scrubbed with wet
cloth;
After removing the stains from surface, use a clean cloth to dip 75% medical alcohol for
further cleaning.
Clean the mechanical surface
All mechanical surfaces of the instrument can be cleaned by wiping with humid cloth. Do not use
any irritative or corrosive cleaning agent. The residual dirt shall be wiped off with the mixture of
50% of normal alcohol and 50% of distilled water and a small amount of household tableware
cleaning liquid.
Disconnect the power supply when wiping the instrument.
Disposal of waste
The wastes generated during use include the sight glass wiping paper or absorbent cotton.
Please don't throw it at will. If there is special garbage disposal facility near you, use it as much
as possible.
The scrapped instruments shall be handled according to the provisions of local environmental
protection laws for avoiding polluting the environment.
Summary of Contents for AM-2000
Page 9: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Safety Caution 5 1 4 3 2 3 5...
Page 11: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Assembly 7 2 1 3 4 5 7 8 6...
Page 14: ...Alltion Guangxi Instrument Co Ltd 10 Product Assembly AM 2000 User s Manual 17 19 20 18...
Page 34: ...Alltion Guangxi Instrument Co Ltd 30 Product Functions AM 2000 User s Manual 1 2 3 4 5 6 7 7 7...
Page 36: ...Alltion Guangxi Instrument Co Ltd 32 Product Functions AM 2000 User s Manual 5 7 3 1 2 4 6 7...