Electromagnetic Compatibility
37
Electromagnetic Compatibility
Without approval from Alltion (Guangxi) Instrument Co., Ltd., it may result in the
electromagnetic compatibility of the device or other equipment if it is not authorized to change
or refit the device.
The design and test of AM-2000 surgical microscope comply with relevant operating
instructions on electromagnetic compatibility.
The equipment or system shall not be adjacent to or stacked together with other equipment. If
it is required, observe and verify whether it can operate correctly in such a configuration.
Requirements on wire installation
Name of Wire
Type
Length (m)
The adapter power cord
Non-shielded parallel wire
3m
Key components for electromagnetic compatibility
The electromagnetic compatibility key components of the product include the power adapter
and dimming circuit board, It will cause significant decreasingly in electromagnetic
compatibility transmission and immunity performance to use or replace with the accessories
with non-matched design,
Do not replace the components without authorization.
Guidance and manufacturers declaration–electromagnetic emission
The AM-2000 surgical microscope is intended for use in the electromagnetic environment
specified below. The customer or the user of the AM-2000 surgical microscope should assure
that it is used in such an environment.
Emissions test
Complianc
e
Electromagnetic Environment - Guidance
RF emissions
CISPR11
Group 1
AM-2000 surgical microscope uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR11
Class A
AM-2000 surgical microscope is suitable for use in al
l establishments, including domestic establishments and t
hose directly connected to the public low-voltage power s
upply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Complies
Summary of Contents for AM-2000
Page 9: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Safety Caution 5 1 4 3 2 3 5...
Page 11: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Assembly 7 2 1 3 4 5 7 8 6...
Page 14: ...Alltion Guangxi Instrument Co Ltd 10 Product Assembly AM 2000 User s Manual 17 19 20 18...
Page 34: ...Alltion Guangxi Instrument Co Ltd 30 Product Functions AM 2000 User s Manual 1 2 3 4 5 6 7 7 7...
Page 36: ...Alltion Guangxi Instrument Co Ltd 32 Product Functions AM 2000 User s Manual 5 7 3 1 2 4 6 7...