21
Communication Select Mode
The Communication Select mode is accessed by pressing the button below “Comm. RS-232” on the
Select Menu display. Please see page 15 for Communication Select mode instructions.
Manufacturer Data Mode
“Manufacturer Data” is accessed by pressing the “Mfg. Data” button on the Select Menu display
(Figure 9, page 14). The “Mfg 1” display shows the name and telephone number of the manufacturer.
The“Mfg 2” display shows important information about your flow meter including the model number,
serial number, and date of manufacture (Figure 11, page 16).
Miscellaneous Mode
The Miscellaneous mode is accessed by pressing the button above the “Misc” label in the upper right
hand corner of the Select Menu display. The screen will appear as shown in Figure 12. Push the
button above “Select” to move the cursor even with the item you wish to adjust. Then use the “UP” and
“DOWN” buttons to make the adjustment. See pages 16 & 17.
NOTE: All Miscellaneous changes are recorded when you exit the Miscellaneous display.
RS-232 Output and Input
Configuring HyperTerminal®:
Open your HyperTerminal® RS-232 terminal program (installed under the “Accessories” menu on
all Microsoft® Windows® operating systems).
Select “Properties” from the file menu.
Click on the “Configure” button under the “Connect To” tab. Be sure the program is set for: 19,200
baud (or matches the baud rate selected in the RS-232 communications menu on the meter) and
an 8-N-1-None (8 Data Bits, No Parity, 1 Stop Bit, and no Flow Control) protocol.
Under the “Settings” tab, make sure the Terminal Emulation is set to ANSI or Auto Detect.
Click on the “ASCII Setup” button and be sure the “Send Line Ends with Line Feeds” box
is not
checked
and the “Echo Typed Characters Locally” box and the “Append Line Feeds to Incoming
Lines” boxes
are checked. Those settings not mentioned here are normally okay in the default
position.
Save the settings, close HyperTerminal® and reopen it.
In Polling Mode, the screen should be blank except the blinking cursor. In order to get the data streaming
to the screen, hit the “Enter” key several times to clear any extraneous information. Type “*@=@”
followed by “Enter” (or using the RS-232 communcation select menu, select @ as identifier and exit the
screen). If data still does not appear, check all the connections and com port assignments.
Changing From Streaming to Polling Mode:
When the meter is in the Streaming Mode, the screen is updated approximately 10-60 times per second
(depending on the amount of data on each line) so that the user sees the data essentially in real time.
It is sometimes desirable, and necessary when using more than one unit on a single RS-232 line, to be
able to poll the unit.
In Polling Mode the unit measures the flow normally, but only sends a line of data when it is “polled”.
Each unit can be given its own unique identifier or address. Unless otherwise specified each unit is
shipped with a default address of capital A. Other valid addresses are B thru Z.
1.
2.
3.
4.
5.
6.