14
Line Pressure
– Pressing the dynamically labeled PSIG button located in the upper left corner of the
display will move the line gauge pressure to the primary display.
Line Temperature
– Pressing the dynamically labeled °C button located in the upper middle of the
display will move the process temperature to the primary display.
Volume
– The volume flow rate is defaulted on the primary display. Pressing the dynamically labeled
“Volume” button will move the volume flow rate to the primary display if another parameter is displayed
there.
Flashing Error Message:
Our flow meters and controllers display an error message (VOV = volumetric
overrange, POV = pressure overrange, TOV = temperature overrange) when a measured parameter
exceeds the range of the sensors in the device. When any item flashes on the display, neither the
flashing parameter nor the flow measurement is accurate. Reducing the value of the flashing parameter
to within specified limits will return the unit to normal operation and accuracy.
Figure 8. Proper Set Up for Remote Tare on Meters (Momentarily ground Pin 4 to Tare)
Select Menu Mode
Pushing “Mode” once will bring up the “Select Menu” display. Push the button nearest your selection to
go to the corresponding screen. Push “Mode” again to return to the Main Mode display. (
Note:
If your
meter was ordered with Totalizing Mode option (page 34), pushing the “Mode” button once will bring up
the “Totalizing Mode” display. Pushing “Mode” a second time will bring up the “Select Menu” display.)
Misc
SELECT
MENU
Comm. Mfg.
RS-232 Data Menu
Figure 9. Select Menu Display