
13
L Series Water Flow Meter Operation
The L Series can have several screen “modes” depending on how the device is ordered. All L Series
meters have a default Main Mode, Select Menu Mode, a Communication Select Mode, a Manufacturer
Data Mode and a Miscellaneous Mode. (In addition, your meter may have been ordered with the
Totalizing Mode option, see page 34.) The device defaults to Main Mode as soon as power is applied
to the meter.
Note
:
L Series Flow Meters may also be powered with the 9 Volt Battery Pack described on page 35.
Main Mode
The main mode screen shows the flow in the units specified at time of order, the line pressure in PSIG,
and the process temperature in degrees Celsius. The following parameters are displayed in the main
mode as shown in Figure 7:
PSIG
o
C Tare
+23.47 +22.73
CCM
Water
+0.000
Volume Main
Figure 7. Main Mode Display, L Series Flow Meter
The “MODE” button in the lower right hand corner toggles the display between modes.
Tare
– Pushing the dynamically labeled Tare button in the upper right hand corner tares the flow meter
and provides it with a reference point for zero flow.
This is a very simple but important step in obtaining
accurate measurements
. It is good practice to “zero” the flow meter each time it is powered up and
whenever a known zero flow condition exists. If the flow reading varies significantly from zero after an
initial tare, give the unit a minute or so to warm up and re-zero it. If possible, it is helpful to zero the unit
near the expected operating pressure by positively blocking
the flow downstream of the flow meter prior
to pushing the “Tare” button. Zeroing the unit while there is
any
flow will directly affect the accuracy by
providing a false zero point. It is not uncommon for a meter to read a small value when it is removed
from the process line as water can “hang” in one or both of the differential pressure sensor ports. If the
unit reads a significant negative value when removed from the line and blocked, it is a good indication
that it was given a false zero.
Note:
A remote tare can be achieved by momentarily grounding pin 4 to tare as shown in Figure 8 on
page 14.