C. Battery Charge and Power On Indicators: Turn green when on
charge. Left indicator turns orange when the A-360 batteries are
below 10% of its capacity.
D. Battery charge connector: Micro-USB Female connector to connect
to PC, battery charger or Docking Station.
5. A-360 BATTERY CHARGE & CHARGE STATUS CHECKING
THE A-360 DOES NOT WORK WHEN IT IS BEING CHARGED.
IT SHOULD BE CHARGED AT LEAST 1.8 M AWAY FROM THE
PATIENT.
The A-360 is supplied with its battery discharged. It is recommended
to perform a full charge/discharge cycle of the battery before starting
to use the A-360 for clinical purposes.
After a full charge, the A-360 can operate, under standard
conditions, for around 120 minutes. Recharging takes
approximately 150 minutes.
Without use, the battery will discharge at a rate of 1% per day.
If the A-360 can be turned on then a minimum of 10 minutes of
operating time is guaranteed.