background image

Air Techniques, Inc.

Page 24

Classic, Ortho & Intraoral

TROUBLESHOOTING

Trouble

Possible Cause

Corrective Action

1 No power - POWER 

switch not lit.

•  Power line cord and or power supply 

adapter not plugged in.

•  No power at Mains Outlet
•  Defective power supply

•  Check that both the line cord and power supply are 

correctly connected.

•  Make sure outlet is grounded and has power.
•  Call your authorized dealer.

2 Blue, Amber or 

Green indicator does 

not light.

•  Defective light or circuitry.

•  Call your authorized dealer.

3 Imaging Software 

does not recog-

nize ScanX when 

selected.

•  Inadequate Computer System.
•  ScanX has not been turned on.

•  The computer connection cable is loose or 

defective.

•  The computer does not recognize that 

ScanX is connected.

•  ScanX hardware problem. 

•  Verify Computer System requirements (Page 10).
•  Make sure that the POWER and ERASER Status LED 

Indicators illuminate blue.

•  Reconnect the cable. Check for tightness. Replace if 

necessary.

•  Verify that the Setup program was correctly 

installed (Page 12).

•  Call your authorized dealer.

4 Plate does not scan 

properly.

•  The PSP was not pushed far enough into 

the ScanX.

•  Worn transport belt or belt driver.

•  Check the plate edges and fully feed the PSP into 

the ScanX.

•  Call your authorized dealer to replace defective 

transport belt or belt driver.

5

No image appears 

after scanning.

•  The PSP fed backwards (printed side 

towards ScanX).

•  The PSP was erased prior to scanning.

•  X-ray source failed or low exposure.
•  Hardware failure.

•  Quickly re-feed the plate with the printed side out. If a 

substandard image results, retake image.

•  Feed the PSPs into the scanner immediately and 

quickly after removal from the cassette.

•  Call your authorized dealer.
•  Call your authorized dealer.

6

Image is too dark.

•  PSP has been overexposed

•  Use software to adjust brightness. If this is not possible, 

retake image with proper (lower) exposure and a newly 

erased PSP.

•  Make sure intensifying screens are removed.

7

Image appears 

skewed on 

monitor.

•  PSP was fed skewed.

•  Worn transport belt or belt driver.

•  When inserting PSP into feed slot, be sure to “feel” 

for resistance of light seal brush, align PSP, and then 

push down uniformly on top edge of PSP.

•  Call your authorized dealer to replace defective 

transport belt or belt driver.

8

Image contains 

ghost images or 

shadows.

•  PSP was not completely erased prior to 

use.

•  Imaging Plate was exposed with the back 

facing the tubehead. 

•  PSP has been stored too long.
•  Partial erasure of the image due to expo-

sure to light during PSP handling. 

•  Make sure ScanX is operating with the eraser turned 

on (blue ERASER LED indicator is lit steadily).

•  Make sure the plates are inserted properly into the 

barrier envelope or cassette with the proper orien-

tation to the X-ray source.

•  Do not store PSPs for more than 24 hours.
•  Do not leave exposed PSPs in well lit areas. Transfer 

PSPs from their protective cassettes or envelope to 

ScanX within one hour of exposure.

9

Image shows 

artifacts or white or 

black lines.

•  The PSP surface is not clean and has dirt, 

stains or scratches on it.

•  ScanX plate transport path may contain 

an obstruction, debris or dust.

•  Clean the PSP with PSP wipes (P/N B8910).

•  Make sure to handle plates properly.

•  Do not reuse the PSP if scratched or stained.
•  Clean transport path using  a ScanX Cleaning Sheet 

(P/N B2010, B2020, B3010 or B3020).

Important:

  Do not allow the 

PSP to be exposed to light 
between X-ray taking and 
scanning with the ScanX .

Summary of Contents for ScanX Classic

Page 1: ...ScanX Classic ScanX Ortho ScanX Intraoral Digital Imaging Systems Instruction Manual ScanX Classic Part No F3700 ScanX Ortho Part No F3750 ScanX Intraoral Part No F3600 ...

Page 2: ...nt for the dental professional since 1962 This manual covers the installation operation and maintenance of 1 ScanX Intraoral Digital Imaging System part number F3600 2 ScanX Classic Digital Imaging System part number F3700 3 ScanX Ortho Digital Imaging System part number F3750 The ScanX Classic and ScanX Ortho are capable of processing both intraoral and extraoral imaging plates while the ScanX In...

Page 3: ...omponent of ScanX is only accessible by removing hardware with special tools and should only be opened and serviced by an authorized dealer service technician Contact your local Air Techniques authorized dealer for service Failure to heed this directive may result in equipment damage and voids the warranty Electrical Safety Notes The power line cord is the mains power disconnect device Use only th...

Page 4: ...hose specified herein may result in hazardous radiation exposure The laser is on only during an active scan Only a trained technician from an authorized dealer should remove a cover from the ScanX Direct eye contact with the output beam from the laser may cause serious damage and possible blindness Equipment Lifting ScanX weighs up to 51 pounds or more and two people may be required to prevent inj...

Page 5: ...EXPOSURE TO BEAM CLASS IIIb LASER PRODUCT COVER REMOVED MAKES THIS DEVICE A ScanX is a Class I Laser Product Class 1 Laser Product IEC This warning label identifies ScanX as such a product and describes the potential danger to humans in the event the product is opened during service There is no laser radiation from this product when operated and maintained as instructed The Laser Product Accession...

Page 6: ...pter B7095 Quantity 1 Quantity 1 Quantity 1 10 Foot Power Cord 73096 Quantity 1 Quantity 1 Quantity 1 6 Foot 2 0 USB Cable B3554 Quantity 1 Quantity 1 Quantity 1 Accessory Kit containing Size 2 Phosphor Storage Plates Quantity 20 Quantity 20 Size 2 Barrier Envelopes Quantity 300 Quantity 300 Size 0 Plate Guide Quantity 1 Quantity 1 Quantity 1 Size 1 Plate Guide Quantity 1 Quantity 1 Quantity 1 Siz...

Page 7: ...again The manufacturer regard themselves as being responsible for the equipment with regard to safety reliability and proper functioning only if assembly resetting changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conformity with the instructions contained in this manual The device conforms to the relevant safety standards valid at t...

Page 8: ...torage hard drive or CD and sharing retrieval An additional feature of ScanX includes a patented in line plate erase function that removes the latent image from the plate immediately after scanning This design provides an efficient one operation scanning and erasing process leaving the user with a PSP ready for the next X ray procedure Features High resolution digital images in seconds Share resul...

Page 9: ...o 80 Non condensing Storage and Transport Temperature 21 F to 130 F 29 C to 54 C Humidity 10 to 80 Non condensing Note Resolution of the device is dependent on operating mode and specific imaging plate type used Resolution LP mm Vertical Horizontal Classic 3 to 18 3 to 12 Intraoral 10 to 18 10 to 12 Compliance Data Laser Classification Class I Laser Product Compliance with FDA HHS 21 CFR 1040 10 a...

Page 10: ...t extended x64 processor Microsoft Windows 7 Professional Enterprise or Ultimate with Service Pack 1 for an Intel 32 bit or an Intel 64 bit extended x64 processor or Microsoft Windows 8 1 Professional or Enterprise for an Intel 32 bit or an Intel 64 bit extended x64 processor USB Port Version USB 2 0 or later Hard Drive 200 MB available disk space required to start scanning Image Management Compat...

Page 11: ... Control Panel and then click the System icon Abbreviations used in this manual are summarized below A ampere s AC alternating current CD ROM compact disk read only memory CFR Code of Federal Regulations CPU central processing unit your computer cm centimeter GB gigabyte 230 109 bytes GHz Gigahertz 109 of Hertz H height Hz Hertz cycles per second IEC International Electro technical Commission IMS ...

Page 12: ...for the first time If not run the Setup program located in the root directory of the CD typically D Autorun exe ScanX Connection Procedure Refer to Figure 2 and perform the following procedure to connect ScanX for operation to a computer for the first time 1 Select a location that meets the Site Selection guidelines 2 Make sure that the computer meets all requirements see page 10 necessary to supp...

Page 13: ...to the slots under of the ScanX 2 Align the two quick connect clips of each support with the corresponding slots 3 Insert each support into the slots until a click is heard 4 Return the ScanX to the upright position on the installation site counter top Detail B Type B USB Connector Type A USB Connector to Computer TO POWER SUPPLY Power Supply Note The power line cord is the Mains disconnect device...

Page 14: ...LED Indicator Displays the erase function status as set by the ERASER Switch 1 Extinguished when the AC MAINS are interrupted No power 2 Extinguished when the system is set to the Standby mode 3 Illuminated blue when the In Line Erase function is enabled Default mode has eraser enabled 4 Flashing blue when the In Line Erase is disabled Numeric Lane Indicator with Visual Pointer LED Indicators Bi C...

Page 15: ... ScanX is set in this mode of operation the POWER and ERASER Status LED Indicators illuminate bright blue as shown by Figure 4 The Numeric Lane Indicator with Visual Pointer LED Indicators are not lit STANDBY Mode of Operation At the end of the day or whenever desired power down ScanX simply by pressing and holding the POWER switch for approximately two seconds until the ERASER and POWER LED Indic...

Page 16: ...X ray exposed and scanned within 24 hours of erasure since natural radiation will add noise to the PSP Erase PSPs by simply using the In Line Erase Feature Erasing of PSPs can be accomplished using one of the following methods Note Both erasing methods will result in an erased PSP suitable for reuse The user will not observe any difference in ScanX operation when using either method Method 1 Perfo...

Page 17: ...pe over the dry Plate surface Wipe back and forth and then in a circular motion 2 Allow the plate surface to air dry Make sure that the PSP is completely dry before re using Disinfecting the Phosphor Storage Plates There is no reason to routinely disinfect the PSPs unless contamination is suspected If a PSP has touched a contaminated surface it may be immersed briefly in a cold sterilant such as a...

Page 18: ...TANT The following procedures are not to be used as a replacement for training in radiography Only trained radiography professionals should perform X ray procedures Notes 1 Theorientationletter a printedonthePSP maybeusedforreference as you would use the dot on an intraoral X ray film In addition a backwards a i e appearing in an image is an indication that the image has been flipped 2 If using ho...

Page 19: ...witch 3 Verify that the POWER and ERASER Status LED Indicators illuminate bright blue 4 Activate the Scanner and select the desired image type and resolution via the imaging software 5 Verify that the four Numeric Lane Indicators with Visual Pointer LED Indicators illuminate bright green as shown by Figure 11 when the Scanner has been activated indicating that the imaging software is set to acquir...

Page 20: ...Plate Guide slot only after the corresponding track indicator LED changes from amber to green 6 Observe that a red glow emanates from the scanner exit slot 7 Repeat steps 2 through 6 to process additional PSPs as necessary Another PSP may be fed into any Plate Guide slot as long as the track indicator light illuminates green 8 Observe that each scanned and erased PSP drops into the receiving tray ...

Page 21: ...Ready mode by pressing the POWER switch 3 Verify that the POWER and ERASER Status LED Indicators illuminate bright blue 4 Activate the Scanner and select the desired image type and resolution via the imaging software 5 Verify that the four Numeric Lane Indicators with Visual Pointer LED Indicators illuminate bright green as shown by Figure 11 when the Scanner has been activated indicating that the...

Page 22: ... 4 Insert the exposed PSPs to be scanned into the ScanX Plate Guides by performing the Scanning and Erasing Plates procedures provided on the previous page When the scanner is operating with the eraser disabled the exit slot will be unlit no red glow 5 The scanned PSPs still contain latent images that require erasure Make sure to erase each PSP prior to reuse for new images Figure 14 Front Panel I...

Page 23: ...h This scratched look on the black side has absolutely no effect on the quality of the image and should be expected under normal conditions If the phosphor side is scratched make sure the plates are being handled properly and not being dragged from ScanX tray area or other surfaces that could cause scratching of the plate Make sure to review the Plate Care and Preparation information provided on p...

Page 24: ...iately and quickly after removal from the cassette Call your authorized dealer Call your authorized dealer 6 Image is too dark PSP has been overexposed Use software to adjust brightness If this is not possible retake image with proper lower exposure and a newly erased PSP Make sure intensifying screens are removed 7 Image appears skewed on monitor PSP was fed skewed Worn transport belt or belt dri...

Page 25: ...r warranty require a Return Materials Authorization RMA Items returned without an RMA or included with other products for which an RMA has been issued may be returned to the customer at the discretion of Air Techniques Any item returned under warranty will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective Air Techniques Inc is ...

Page 26: ...248 3 100 4 73248 4 50 ScanX Cleaning Sheets Description Order No Qty ScanX Classic Sheets B2010 12 ScanX Classic Sheets B2020 25 ScanX Intraoral Sheets B3010 12 ScanX Intraoral Sheets B3020 25 One re usable Release Liner is included in each box of Sheets Intraoral Phosphor Storage Plates Size Order No Qty 0 73445 0 2 1 73445 1 2 2 73445 2 4 3 73445 3 2 4 73445 4 1 Extraoral Phosphor Storage Plate...

Page 27: ...mless integration with all our digital imaging ScanX Systems and Polaris Intraoral Cameras Key Features Easy to learn and simple to use Acquire mount view and store images with just a few mouse clicks Customize images to personal preferences with our simple Set Up 1 2 3 tool Exposure Our Stop Light Exposure System is a unique indicator to ensure optimum data capture It helps you acquire great imag...

Page 28: ...turing reliable products and providing excellent customer and technical support Whether the need is digital imaging utility room equipment or merchandise Air Techniques can provide the solution via our network of authorized professional dealers Proudly designed tested and manufactured in the U S our products are helping dental professionals take their practices to the next level Air Techniques fam...

Reviews: