Page 3

Air Techniques, Inc. 

Classic, Ortho & Intraoral

ScanX has been designed to minimize exposure of personnel to hazards . While the equipment is designed for 
safe operation, certain precautions must be observed .  Use of ScanX 

not

 in conformance with the instructions 

specified in this manual may result in permanent failure of the unit .

General

.

• Check with your dealer for packing material requirements if it is necessary to return the product to 

the manufacturer . Correct packing guarantees optimal safety of the device during transport . Should it 
become necessary to return the device to the manufacturer during the warranty period or other times, 
the manufacturer will not accept claims for damage arising from using incorrect packing materials . 

• Before every use, the operator must check the functional safety and the condition of the device .
• The operator must be knowledgeable in the operation of the device . 

Use of Accessory Equipment.

The use of ACCESSORY equipment not complying with the equivalent safety requirements of this 
equipment may lead to a reduced level of safety of the resulting system . 
Use of ACCESSORIES or cables other than those specified or provided by the manufacturer may result 
in increased EMISSIONS or decreased IMMUNITY of the EQUIPMENT .

Do Not Attempt Internal Service.

The interior of each component of ScanX is only accessible by removing hardware with special tools 
and should only be opened and serviced by an authorized dealer service technician .
Contact your local Air Techniques authorized dealer for service . Failure to heed this directive may result in 
equipment damage and voids the warranty .

Electrical Safety Notes.

  The power line cord is the mains power disconnect device .

  Use only the line cord and power supply provided with the unit .

  Use only grounded electrical connections .

  To avoid risk of electric shock, fire, short-circuit or dangerous emissions, never insert 

any metallic object into the equipment .

  Only use connection cable(s) delivered with the device .

  Check the device cables for possible damage before switching on . Damaged cables, plugs 

and sockets must be replaced before use .

  Never touch open supply outlets and patients simultaneously .

  Do not locate unit where it could be sprayed with water, or in a damp environment .

GENERAL SAFETY

This document is a guide to the proper use of ScanX .  It provides the information necessary for the setup, 
operation and routine care and maintenance of the device .  Review and follow the guidelines included in 
this Instruction Manual to ensure that your ScanX gives you the highest level of performance .
This manual is not to be used as a replacement for training in radiography .
For information regarding the computer system and imaging software, refer to the appropriate 
documentation provided with your computer hardware and software .

PURPOSE OF THIS MANUAL

Summary of Contents for ScanX Classic

Page 1: ...ScanX Classic ScanX Ortho ScanX Intraoral Digital Imaging Systems Instruction Manual ScanX Classic Part No F3700 ScanX Ortho Part No F3750 ScanX Intraoral Part No F3600 ...

Page 2: ...nt for the dental professional since 1962 This manual covers the installation operation and maintenance of 1 ScanX Intraoral Digital Imaging System part number F3600 2 ScanX Classic Digital Imaging System part number F3700 3 ScanX Ortho Digital Imaging System part number F3750 The ScanX Classic and ScanX Ortho are capable of processing both intraoral and extraoral imaging plates while the ScanX In...

Page 3: ...omponent of ScanX is only accessible by removing hardware with special tools and should only be opened and serviced by an authorized dealer service technician Contact your local Air Techniques authorized dealer for service Failure to heed this directive may result in equipment damage and voids the warranty Electrical Safety Notes The power line cord is the mains power disconnect device Use only th...

Page 4: ...hose specified herein may result in hazardous radiation exposure The laser is on only during an active scan Only a trained technician from an authorized dealer should remove a cover from the ScanX Direct eye contact with the output beam from the laser may cause serious damage and possible blindness Equipment Lifting ScanX weighs up to 51 pounds or more and two people may be required to prevent inj...

Page 5: ...EXPOSURE TO BEAM CLASS IIIb LASER PRODUCT COVER REMOVED MAKES THIS DEVICE A ScanX is a Class I Laser Product Class 1 Laser Product IEC This warning label identifies ScanX as such a product and describes the potential danger to humans in the event the product is opened during service There is no laser radiation from this product when operated and maintained as instructed The Laser Product Accession...

Page 6: ...pter B7095 Quantity 1 Quantity 1 Quantity 1 10 Foot Power Cord 73096 Quantity 1 Quantity 1 Quantity 1 6 Foot 2 0 USB Cable B3554 Quantity 1 Quantity 1 Quantity 1 Accessory Kit containing Size 2 Phosphor Storage Plates Quantity 20 Quantity 20 Size 2 Barrier Envelopes Quantity 300 Quantity 300 Size 0 Plate Guide Quantity 1 Quantity 1 Quantity 1 Size 1 Plate Guide Quantity 1 Quantity 1 Quantity 1 Siz...

Page 7: ...again The manufacturer regard themselves as being responsible for the equipment with regard to safety reliability and proper functioning only if assembly resetting changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conformity with the instructions contained in this manual The device conforms to the relevant safety standards valid at t...

Page 8: ...torage hard drive or CD and sharing retrieval An additional feature of ScanX includes a patented in line plate erase function that removes the latent image from the plate immediately after scanning This design provides an efficient one operation scanning and erasing process leaving the user with a PSP ready for the next X ray procedure Features High resolution digital images in seconds Share resul...

Page 9: ...o 80 Non condensing Storage and Transport Temperature 21 F to 130 F 29 C to 54 C Humidity 10 to 80 Non condensing Note Resolution of the device is dependent on operating mode and specific imaging plate type used Resolution LP mm Vertical Horizontal Classic 3 to 18 3 to 12 Intraoral 10 to 18 10 to 12 Compliance Data Laser Classification Class I Laser Product Compliance with FDA HHS 21 CFR 1040 10 a...

Page 10: ...t extended x64 processor Microsoft Windows 7 Professional Enterprise or Ultimate with Service Pack 1 for an Intel 32 bit or an Intel 64 bit extended x64 processor or Microsoft Windows 8 1 Professional or Enterprise for an Intel 32 bit or an Intel 64 bit extended x64 processor USB Port Version USB 2 0 or later Hard Drive 200 MB available disk space required to start scanning Image Management Compat...

Page 11: ... Control Panel and then click the System icon Abbreviations used in this manual are summarized below A ampere s AC alternating current CD ROM compact disk read only memory CFR Code of Federal Regulations CPU central processing unit your computer cm centimeter GB gigabyte 230 109 bytes GHz Gigahertz 109 of Hertz H height Hz Hertz cycles per second IEC International Electro technical Commission IMS ...

Page 12: ...for the first time If not run the Setup program located in the root directory of the CD typically D Autorun exe ScanX Connection Procedure Refer to Figure 2 and perform the following procedure to connect ScanX for operation to a computer for the first time 1 Select a location that meets the Site Selection guidelines 2 Make sure that the computer meets all requirements see page 10 necessary to supp...

Page 13: ...to the slots under of the ScanX 2 Align the two quick connect clips of each support with the corresponding slots 3 Insert each support into the slots until a click is heard 4 Return the ScanX to the upright position on the installation site counter top Detail B Type B USB Connector Type A USB Connector to Computer TO POWER SUPPLY Power Supply Note The power line cord is the Mains disconnect device...

Page 14: ...LED Indicator Displays the erase function status as set by the ERASER Switch 1 Extinguished when the AC MAINS are interrupted No power 2 Extinguished when the system is set to the Standby mode 3 Illuminated blue when the In Line Erase function is enabled Default mode has eraser enabled 4 Flashing blue when the In Line Erase is disabled Numeric Lane Indicator with Visual Pointer LED Indicators Bi C...

Page 15: ... ScanX is set in this mode of operation the POWER and ERASER Status LED Indicators illuminate bright blue as shown by Figure 4 The Numeric Lane Indicator with Visual Pointer LED Indicators are not lit STANDBY Mode of Operation At the end of the day or whenever desired power down ScanX simply by pressing and holding the POWER switch for approximately two seconds until the ERASER and POWER LED Indic...

Page 16: ...X ray exposed and scanned within 24 hours of erasure since natural radiation will add noise to the PSP Erase PSPs by simply using the In Line Erase Feature Erasing of PSPs can be accomplished using one of the following methods Note Both erasing methods will result in an erased PSP suitable for reuse The user will not observe any difference in ScanX operation when using either method Method 1 Perfo...

Page 17: ...pe over the dry Plate surface Wipe back and forth and then in a circular motion 2 Allow the plate surface to air dry Make sure that the PSP is completely dry before re using Disinfecting the Phosphor Storage Plates There is no reason to routinely disinfect the PSPs unless contamination is suspected If a PSP has touched a contaminated surface it may be immersed briefly in a cold sterilant such as a...

Page 18: ...TANT The following procedures are not to be used as a replacement for training in radiography Only trained radiography professionals should perform X ray procedures Notes 1 Theorientationletter a printedonthePSP maybeusedforreference as you would use the dot on an intraoral X ray film In addition a backwards a i e appearing in an image is an indication that the image has been flipped 2 If using ho...

Page 19: ...witch 3 Verify that the POWER and ERASER Status LED Indicators illuminate bright blue 4 Activate the Scanner and select the desired image type and resolution via the imaging software 5 Verify that the four Numeric Lane Indicators with Visual Pointer LED Indicators illuminate bright green as shown by Figure 11 when the Scanner has been activated indicating that the imaging software is set to acquir...

Page 20: ...Plate Guide slot only after the corresponding track indicator LED changes from amber to green 6 Observe that a red glow emanates from the scanner exit slot 7 Repeat steps 2 through 6 to process additional PSPs as necessary Another PSP may be fed into any Plate Guide slot as long as the track indicator light illuminates green 8 Observe that each scanned and erased PSP drops into the receiving tray ...

Page 21: ...Ready mode by pressing the POWER switch 3 Verify that the POWER and ERASER Status LED Indicators illuminate bright blue 4 Activate the Scanner and select the desired image type and resolution via the imaging software 5 Verify that the four Numeric Lane Indicators with Visual Pointer LED Indicators illuminate bright green as shown by Figure 11 when the Scanner has been activated indicating that the...

Page 22: ... 4 Insert the exposed PSPs to be scanned into the ScanX Plate Guides by performing the Scanning and Erasing Plates procedures provided on the previous page When the scanner is operating with the eraser disabled the exit slot will be unlit no red glow 5 The scanned PSPs still contain latent images that require erasure Make sure to erase each PSP prior to reuse for new images Figure 14 Front Panel I...

Page 23: ...h This scratched look on the black side has absolutely no effect on the quality of the image and should be expected under normal conditions If the phosphor side is scratched make sure the plates are being handled properly and not being dragged from ScanX tray area or other surfaces that could cause scratching of the plate Make sure to review the Plate Care and Preparation information provided on p...

Page 24: ...iately and quickly after removal from the cassette Call your authorized dealer Call your authorized dealer 6 Image is too dark PSP has been overexposed Use software to adjust brightness If this is not possible retake image with proper lower exposure and a newly erased PSP Make sure intensifying screens are removed 7 Image appears skewed on monitor PSP was fed skewed Worn transport belt or belt dri...

Page 25: ...r warranty require a Return Materials Authorization RMA Items returned without an RMA or included with other products for which an RMA has been issued may be returned to the customer at the discretion of Air Techniques Any item returned under warranty will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective Air Techniques Inc is ...

Page 26: ...248 3 100 4 73248 4 50 ScanX Cleaning Sheets Description Order No Qty ScanX Classic Sheets B2010 12 ScanX Classic Sheets B2020 25 ScanX Intraoral Sheets B3010 12 ScanX Intraoral Sheets B3020 25 One re usable Release Liner is included in each box of Sheets Intraoral Phosphor Storage Plates Size Order No Qty 0 73445 0 2 1 73445 1 2 2 73445 2 4 3 73445 3 2 4 73445 4 1 Extraoral Phosphor Storage Plate...

Page 27: ...mless integration with all our digital imaging ScanX Systems and Polaris Intraoral Cameras Key Features Easy to learn and simple to use Acquire mount view and store images with just a few mouse clicks Customize images to personal preferences with our simple Set Up 1 2 3 tool Exposure Our Stop Light Exposure System is a unique indicator to ensure optimum data capture It helps you acquire great imag...

Page 28: ...turing reliable products and providing excellent customer and technical support Whether the need is digital imaging utility room equipment or merchandise Air Techniques can provide the solution via our network of authorized professional dealers Proudly designed tested and manufactured in the U S our products are helping dental professionals take their practices to the next level Air Techniques fam...

Reviews: