Note: When adjusting the target value in the main interface, press
the trigger, and the wrench will rotate.
d) Run the tool on the test stand and record the torque result. Once
operation is finished, press key
to return to where you left off in the
Sixpoint menu. Press
key
to scroll to the Result column to the right
of the Target value.
Press key
to select. This will open the value
setting screen where you can input the torque result.
Scrolls
forward
Scrolls backward
Key
1
Key
2
Key
3
Key
2
Selects
highlighted digit
and opens value
setting screen
Selects highlighted
number
Key
2
When
finished,
select
OK then
to return
to the
Sixpoint
value
screen
e) Press key
or
to scroll through the value table. Press key
to
select the highlighted number. The number will appear at the top of
the screen.
Follow the same steps for the remaining digits, pressing key
to
select each digit. If you make a mistake or need to modify the set
value, select “-”, and press key
to delete one digit.
When the value is completed, move to “OK”, and press key
to
complete the setting. This returns to the Sixpoint value screen.
f) Input the remaining 5 calibration value sets using the same process.
g) When finished setting all values, select OK using key
. The system
will automatically calibrate and will return you to the Calibration screen.
Parameter
a) From the Calibration menu, press key
or
to navigate to
Parameter and press key
to select.