18 - Specification
Digital Input/Output (DIO)
98
SMU4000 Series Instruction Manual
Digital Input/Output (DIO)
Signals
5 user definable signal pins
HV interlock (SMU4201 only)
Global Trigger In
Global Trigger out (self-resetting based on trigger input)
Ground and +5V (fuse limited to 500mA)
Voltage Reference
Chassis ground referenced
Input Levels
Logic zero:
-0.25V to +1V (diode clamped to 0V)
Logic one:
+1.75V to +5.25V (diode clamped to +5V)
Output Levels
Logic zero:
open-drain MOS, typically 0.2ohm, 100mA
maximum sink
Logic one:
nominally 4.7kohm pull-up to +5V
Logic
Active High / Low global port selection
Remote Control Interfaces
LAN
Ethernet 100/10 base-T hardware connection.
LXI Conformance
1.5 LXI Device Specification 2016
USB
Standard USB 2.0 hardware connection. Operates as a virtual
COM port.
GPIB (optional)
Conforming with IEEE488.1 and IEEE488.2
Remote Command
Processing Time
Typically < 5ms between receiving the command terminator
for an output change at the instrument and the output
beginning to change.
System Speeds
Measure to Memory
Up to 5k/s sustained