16 - Calibration
Measurement Zero and Self Calibration
90
SMU4000 Series Instruction Manual
CALIBRATION
Measurement Zero and Self Calibration
An automatic zero calibration and self calibration is performed every time that the instrument is
switched on.
However, if the instrument has been stored at a temperature outside the specified operating
range, and is switched on before it has fully acclimatised to the working environment, accuracy
may be affected as the meter’s temperature changes.
To ensure optimum accuracy, particularly on the lower current and voltage ranges, zero
calibration can be repeated when the meter has acclimatised by using the TEST key on the front
panel.
NOTE
It is recommended that the self test is performed after 1 hour warm up or before acquiring
measurements for highly accurate tests .
Routine Calibration
The Calibration menu provides access to all of the instrument calibration routines.
The external calibration routines require accurate measurement equipment for measuring ultra-
precise DC voltages and currents.
This menu is password protected; the password can be found in the SMU4000 Series Service
Guide, available upon request from
WARNING
Routine Calibration should only be carried out by qualified service personnel. Some of the
calibration procedures may expose the service personnel to hazardous voltages.