2 - Safety
Safety notices
6
SMU4000 Series Instruction Manual
Safety notices
This instrument is:
·
A safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010-1 and EN61010-2-030 (Safety
Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use).
·
An Installation Category II instrument intended for operation from a normal
single-phase supply.
·
Supplied in a safe condition and tested in accordance with EN61010-1.
·
Designed for indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 40°C, 20% - 80% RH (non-condensing) and less than
2000m altitude.
WARNING
Do not operate while condensation is present.
Do not operate outside its rated supply voltages or environmental range.
THIS INSTRUMENT MUST BE EARTHED.
Any interruption of the mains earth connector, inside or outside, will make the instrument
dangerous. Intentional interruption is prohibited.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided
If any adjustment or repair of the opened power supply under voltage is inevitable it shall be
carried out only by a skilled person who is trained to perform such adjustments and is aware of
the hazards involved.
When connected, terminals may be live and opening the covers or removal of parts (except
those that can be accessed by hand) may expose live parts.
To avoid electric shock, or damage to the instrument, never allow water to get inside the case.
If the instrument is clearly defective, has been subject to mechanical damage, excessive
moisture, or chemical corrosion the safety protection may be impaired and it must be
withdrawn from use and returned for checking and repair.
Voltages above 60VDC are hazardous live according to EN 61010-1 and great care must be taken
when using the SMU at voltages above this level.
Capacitors inside the SMU may still be charged even if it has been disconnected from all voltage
sources, these will be safely discharged a few minutes after switching off. LEDs on the HV rail to
indicate that charge is still present, as such it’s not safe to dismantle until all LEDs have gone
out.
This instrument is protected by three internal fuses which are user serviceable (refer to the
Service Manual).
CAUTION
Do not wet when cleaning, use only a soft dry cloth to clean the screen.
Do not use sharp or pointed objects to operate the touch screen.
Take care not to restrict the inlet vents at the front, rear, sides or underneath the instrument. In
rack-mounted situations, allow adequate space around the instrument.