6 - Getting Started
Connections
19
SMU4000 Series Instruction Manual
GETTING STARTED
Connections
Connections can be made to the front or rear terminals on the instrument. The front and rear
terminals are connected in parallel and must
never
be used simultaneously.
Front/Rear connections
Front and rear terminals are always active, there is no need to select between the two.
The terminals on the front panel are 4mm banana jack connectors, and the rear-panel contains
a terminal block. The rear terminals offer additional Guard connections for screening, see
’ for more details.
Front panel Main HI and LO terminals are permanently wired in parallel with the rear panel F +/-
[Force] terminals.
Front panel SENSE HI and LO terminals are permanently wired in parallel with the rear panel S
+/- [Sense] terminals
WARNING
Only ever use either the front or rear terminals, never use both at once.
Hazardous
voltages
will appear on both sets of terminals.
Hazardous voltages will appear on both sets of terminals
.
Always make connections to the instrument with the OUTPUT off
and in the Open off state;
this is the only output state that completely isolates the external circuitry from the
instrument.