12 - Application notes
Source- Measure considerations
81
SMU4000 Series Instruction Manual
APPLICATION NOTES
Source- Measure considerations
Stability
Due to their fast and agile output capabilities, source measure units are sensitive to the impact
of load characteristics, interconnection inductance and feedback loop characteristics, which can
give rise to unexpected instability or poor dynamic behaviour. However, there are key steps that
can be taken to help over-come these potential issues.
•
Ensure all current carrying output wires (MAIN HI and MAIN LO) are as short and thick
as possible.
•
Ensure all current carrying output wires (MAIN HI and MAIN LO) are twisted tightly
together to minimise loop area.
•
Enable high reactance mode.
•
Fix voltage and current ranges to a suitable manual selection.
•
Be very cautious when utilizing the 200nA and 2µA ranges, as these ranges are very
sensitive. Try selecting the 20µA range instead.
•
Reduce the maximum slew rate.
•
Reduce all output load inductance and capacitance as much as possible.
Noise
Pickup
Due to their high resolution and numerous low level measurement ranges, source measure
units are very susceptible to noise pickup. However, there are many things that can be done to
help reduce noise pickup issues.
•
Always set an integer number of PLCs for the measurement period. Extend the
measurement period as long as possible. It’s recommended to use at least 50 PLC for
the 200nA range.
•
Utilize the guard capability. Ensure the guard is kept as close to the Main Hi signal as
possible. This can significantly reduce leakage issues.
•
When utilizing the 2-wire terminal configuration, ensure the cables are twisted tightly
together. Essentially shield the MAIN HI signal with the MAIN LO signal. Alternatively
use a 4mm banana to BNC adaptor and use a screened BNC cable instead. Ensure the
BNC cable is correctly rated for the voltage and current levels used.
•
Where possible try to place the DUT into a shielded enclosure that is connected to
MAIN LO.
•
Utilize the front panel ground chassis connection as a shield, however this is unlikely to
be as effective as the MAIN LO terminal.
•
Turn off/ route cables away from any known noise sources within the environment.