3 - Installation
Electrical Requirements
8
SMU4000 Series Instruction Manual
Electrical Requirements
Mains operating voltage
CAUTION
The operating voltage is switch selectable between 115Vac or 230Vac. You must check that the
local supply meets the AC input printed on the rear panel before connecting the unit to the
supply.
Changing the mains operating voltage
Should it be necessary to change the operating voltage from 230V to 115V or vice-versa, follow
the instructions below:
WARNING
The apparatus must be disconnected from all voltage sources before it is opened for any
adjustment. Capacitors inside the SMU may still be charged even if it has been disconnected
from all voltage sources, these will be safely discharged a few minutes after switching off. LEDs
on the HV rail to indicate that charge is still present, as such it’s not safe to dismantle until all
LEDs have gone out.
Remove the case:
Mains supply voltage selection:
①
Position the handle as shown.
③
Slide and lift the case clear.
②
Remove 2x screws from the rear panel.
A switch is located in the highlighted postiton (SW1).
Select the required value with the switch
and replace the case.
CAUTION
The set AC input voltage MUST be clearly
labeled on the rear panel.